10.3 Battery replacement

Warning : change the battery only in a non hazardous
area.
• Unscrew the 2 screws of the battery holder using the 2,5 mm
Hex Allen key which is located on the carrying case. See
Figure 10-4.
one-way only device
• Pull it gently out.
• Change the battery (one-way only device). See Figure 10-4.
• Push the battery holder back in its housing (one-way
only).
• Tighten the 2 screws.
Only one battery is approved:
Duracell / Procell MN1604
one-way only device
Figure 10-4