Required Equipment; Accuracy Test - Fluke 28 II Ex Calibration Information Manual

Digital multimeter
Hide thumbs Also See for 28 II Ex:
Table of Contents

Advertisement

Required Equipment

Table 3 is a list of the equipment necessary to do performance tests on the Meter.
Recommended Equipment
Calibrator
(Fluke 5520A or equivalent)
K-type Thermocouple, mini-plug
on both ends

Accuracy Test

To measure the accuracy of the Meter, set the Calibrator to the Calibrator output
parameters shown in Table 4. Make sure the UUT measurement is between the low and
high limit shown in the table.
Test
Step
Range
Function
K
1
600 mV
AC Volts
2
600 mV
3
600 mV
4
6 V
5
6 V
6
6 V
7
60 V
8
60 V
9
600 V
10
600 V
11
600 V
12
1000 V
13
1000 V
Table 3. Required Equipment
Measurement Function
DC Volts
DC Current
AC Volts
AC Current
Resistance
Capacitance
Frequency
Temperature
Table 4. Accuracy Tests
Calibrator Output
60 mV, 60 Hz
330 mV, 60 Hz
600 mV, 13 kHz
675 mV, 60 Hz
3.3 V, 60 Hz
3.3 V, 20 kHz
6.75 V, 60 Hz
33 V, 60 Hz
67.5 V, 60 Hz
330 V, 60 Hz
330 V, 2.5 kHz
100 V, 60 Hz
1000 V, 1 kHz
Accuracy
0 V to 1000 V ±0.012 %
350 μA to 2 A ±0.05 %
0 V to 1000 V ±0.15 %
@ 60 Hz to 20 kHz ±3 %
350 μA to 2 A ±0.39 %
@ 60 Hz to 1 kHz
1 Ω to 100 MΩ ±0.06 %
9 μF to 900 μF ±0.475 %
19.999 kHz to 199.99 kHz,
±0.0137 %
150 mV to 6 Vrms, ±5 %
UUT Measurement
Low Limit
High Limit
59.2 mV
60.8 mV
327.3 mV
332.7 mV
586.0 mV
614.0 mV
0.666 mV
0.684 mV
3.273 V
3.327 V
3.214 V
3.386 V
6.68 V
6.82 V
32.75 V
33.25 V
66.8 V
68.2 V
327.5 V
332.5 V
323.0 V
337.0 V
97 V
103 V
986 V
1014 V
Digital Multimeter
Performance Tests
13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents