Calibration Adjustments - Fluke 28 II Ex Calibration Information Manual

Digital multimeter
Hide thumbs Also See for 28 II Ex:
Table of Contents

Advertisement

Test
Step
Function
42
DC Microamps
43
44
Capacitance
45
46
47
K
Low Pass
48
Filter
49
VDC Peak
Min/Max
50
mVdc
Temperature
[4]
51
[1] Remove test leads from unit.
[2] Use REL to compensate for internal Meter and lead capacitance (must disconnect test leads from calibrator before pushing REL)
[3] The Meter accuracy is not specified at this input signal frequency with Low-pass filter selected. The display reading shown, checks
that the Low-pass filter is active and follows an expected roll-off curve.
[4] To ensure accurate measurement, the Meter and thermocouple adapter must be at the same temperature. After connecting the
thermocouple adapter to the Meter allow for reading to stabilize before recording display reading.
Backlight Functional Test
A backlight test is done to make sure the backlight comes on with the first push of H.
The second push causes the backlight to be brighter and a third push turns off the
backlight.

Calibration Adjustments

If one or more of the accuracy tests shows a measurement that is not between the high
and low limits, you can do adjustments. This adjustment procedure sets the Meter to
operate to its specifications.
If the adjustment routine is stopped before you complete the procedure, no
changes are made to the calibration constants in memory.
You must set the Meter to the CAL mode and type in a password to do calibration
adjustments.
Cal Mode Button Functions
For the CAL mode, some of the buttons on the Meter have alternative functions. Some
buttons function differently at different points in the calibration adjustment procedure.
Table 5 is a list of the Meter buttons and their function for the calibration adjustment
procedure.
Table 4. Accuracy Tests (cont.)
Range
5520A Output
600 μA
330 μA
6000 μA
3300 μA
[1]
10 nF
Open input
100 nF
5 nF
[2]
100 μF
9.5 μF
1000 V
400 V, 400 Hz
1000 V
400 V, 800 Hz
6 Vpp, 500 Hz Sq.
6 V
Wave, DC offset 1 V
0 °C
100 °C
Note
UUT Measurement
Low Limit
328.9 μA
3291 μA
0.70 nF
4.7 nF
9.2 μF
372 V
[3]
226 V
-1.798 V to -2.202 V
-1.0 °C
98.0 °C
Digital Multimeter
Calibration Adjustments
High Limit
331.1 μA
3309 μA
1.10 nF
5.3 nF
9.8 μF
408 V
340 V
3.797 V to 4.203 V
1.0 °C
102.0 °C