Cytometer Control Parameter Setup

This screen is used to select the Signals required in the protocol. The signals that are available are
selectable by checking the options you require.
NOTE
If you add a parameter to an existing protocol, the Detector Names revert to the default names shown
below. You cannot make any changes on this screen when running a
CAUTION
Risk of erroneous results if you assign the same parameter name to more than one
detector. Use a different parameter name for each detector.
You can arrange the order within the
item(s) and Drag and Drop into your chosen order within the
CAUTION
Risk of erroneous results if you modify a protocol by removing or changing the
order of selected parameters. Verify that the protocol's plots, regions and gate
assignments are as desired before reporting results.
To remove a previously selected item from the
longer required.
Ratio
– Choose from the drop down box the appropriate option for the Numerator and Denominator
if required.
A detector name may be entered, this is displayed in all
appear on plots.
773232AH
Selected Signals
list box by highlighting the appropriate
Selected Signals
Selected Signals
Cytometer Control
Using Navios Software
Cytometer Menu
locked
protocol.
list box.
box, uncheck the parameter no
screens but does not
10
10-163