Sample Period

The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measure-
ments transmitted from the serial interface. If the sample rate is set to 5, the
bath transmits the current measurement over the serial interface approximately
every five seconds. The automatic sampling is disabled with a sample period of
0. The sample period is indicated by,
Serial sample period parameter
SAmPLE
Press "SET" to choose to set the sample period. The current sample period
value will be displayed.
Current sample period (seconds)
SA= 1
Adjust the value with "UP" or "DOWN" and then use "SET" to set the sample
rate to the displayed value.
New sample period
SA= 60

Duplex Mode

The next parameter is the duplex mode. The duplex mode may be set to full du-
plex or half duplex. With full duplex any commands received by the bath via
the serial interface are immediately echoed or transmitted back to the device of
origin. With half duplex the commands are executed but not echoed. The du-
plex mode parameter is indicated by,
Serial duplex mode parameter
dUPL
Press "SET" to access the mode setting.
Current duplex mode setting
dUP=FULL
The mode may be changed using "UP" or "DOWN" and pressing "SET".
New duplex mode setting
dUP=HALF

Linefeed

The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
Serial linefeed parameter
LF