Specifications And Environmental Conditions; Specifications - Fluke 7012 User Manual

Calibration bath
Table of Contents

Advertisement

3
Specifications and Environmental
Conditions
3.1

Specifications

Range
Stability
Uniformity
Temperature
Setting
Set-point
Resolution
Display
Resolution
Digital Setting
Accuracy
Digital Setting
Repeatability
Heaters
Safety Cutout
Temperature
Drift per ±°C
Ambient or ±10
VAC line
Pump
Recirculation
(optional)
Access Opening
Depth
Wetted Parts
Volume
Weight
Size
Power
System Fuse

3 Specifications and Environmental Conditions

–10°C to 110°C (14°F to 230°F)
±0.0008°C at 0°C (ethanol)
±0.0008°C at 25°C (water)
±0.003°C at 100°C (oil 5012)
±0.003°C at 0°C (ethanol)
±0.002°C at 25°C (water)
±0.004°C at 100°C (oil 5012)
Digital display with push-button data entry
0.002°C; high-resolution mode, 0.00003°C
0.01°C
±1°C
±0.005°C
500 and 1000 Watts
Factory-set high temperature
±0.2m°C
15 L/min
162 mm X 292 mm (6.38" X 11.5")
457 mm (18")
304 stainless steel
42 liters (11.2 gallons)
68 kg (150 lb.)
762 mm H X 686 mm W X 401 mm D
(30" X 27" X 15.8")
115 VAC (±10%), 60 Hz, 1570 W
230 VAC (±10%), 60 Hz, 1570 W (50 Hz Optional)
NOTE: If voltage is outside ±10% VAC, the compressor may be damaged.
Check the back panel label for the correct voltage and frequency prior to energiz-
ing the instrument.
115 VAC: 15 A 250 V slow blow
230 VAC: 8 A 250 V slow blow