7012 Calibration Bath
User's Guide
DO NOT change the values of the bath calibration constants from the factory set
9.14.1
9.14.2
50
values. The correct setting of these parameters is important to the safety
and proper operation of the bath.
The calibration parameters menu is indicated by,
Calibration parameters menu
CAL
Press "SET" five times to enter the menu.

CTO

Parameter CTO sets the calibration of the over-temperature cutout. This is not
adjustable by software but is adjusted with an internal potentiometer. For the
7012 baths this parameter should read between 120.

H and L

These parameters set the upper and lower set-point limits of the bath. DO NOT
change the values of these parameters from the factory set values. To do so may
present danger of the bath exceeding its temperature range causing damage or
fire.