Press "SET" to access the linefeed parameter.
LF= On
The mode may be changed using "UP" or "DOWN" and pressing "SET".
LF= OFF
9.13

IEEE-488 Parameters

Baths may optionally be fitted with an IEEE-488 GPIB interface. In this case
the user may set the interface address within the IEEE-488 parameter menu.
This menu does not appear on baths not fitted with the interface. The menu is
indicated by,
Press "SET" to enter the menu.
9.13.1

IEEE-488 Address

The IEEE-488 interface must be configured to use the same address as the ex-
ternal communicating device. The address is indicated by,
AddrESS
Press "SET" to access the address setting.
Add= 22
Adjust the value with "UP" or "DOWN" and then use "SET" to set the address
to the displayed value.
Add= 15
9.14

Calibration Parameters

The operator of the bath controller has access to a number of the bath calibra-
tion constants namely CTO, H, and L. These values are set at the factory and
must not be altered. The correct values are important to the accuracy and
proper and safe operation of the bath. Access to these parameters is available to
the user only so that in the event that the controller's memory fails the user may
restore these values to the factory settings. The user should have a list of these
constants and their settings with the manual.
Current linefeed setting
New linefeed setting
IEEE-488 parameters menu
IEEE
I EEE-488 interface address
Current IEEE-488 interface address
New IEEE-488 interface address
9 Controller Operation
IEEE-488 Parameters
49