Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The BG5S is intended for use in the electromagnetic environment specified below. The customer or the user
of the BG5S should assure that it is used in such an environment.
IMMUNITY test
IMMUNITY test
Conducted RF
3 Vrms 150 kHz
IEC 61000-4-6
to 80 MHz
Radiated RF
3 V/m 80 MHz to
IEC 61000-4-3
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radiOS , amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
BG5S is used exceeds the applicable RF compliance level above, the BG5S should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the BG5S.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
IMMUNIT
Y test
Portable
equipment should be used no closer to any part of
the
BG5S,
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance: : : :
3 Vrms
d
1
2 .
=
1
2 .
3 V/m
d
=
2
3 .
d
=
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment
marked with the following symbol: