Electromagnetic Compatibility Information - iHealth BG5S Owner's Manual

Wireless smart gluco-monitoring system
Hide thumbs Also See for BG5S:
Table of Contents

Advertisement

This product is approved in accordance to RED directive.

ELECTROMAGNETIC COMPATIBILITY INFORMATION

The BG5S is intended for use in the electromagnetic environment specified below.
The customer or the user of the BG5S should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
The BG5S is intended for use in the electromagnetic environment specified below. The customer or the user
of the BG5S should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Table 1 For All ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic emissions
Compliance
The BG5S uses RF energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The BG5S is suitable for use in all establishments, including
Class B
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
Class A
domestic purposes.
Complies
Table 2 For All ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601test
level
± 6 kV contact
± 8 kV air
±2kV for power
supply lines
±1kV for
input/output lines
3 A/m
Electromagnetic environment - guidance
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
±1kV for input/output
lines
3 A/m
26 / 27
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic
material,
humidity should be at least 30 %.
Mains power quality should be that of
a typical commercial or hospital
environment.
Power
frequency
magnetic
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
the
relative
fields