Connecting The Siphon Overflow - Agilent Technologies 1260 Infinity User Manual

Infinity evaporative light scattering detector
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

3
Installing the Detector
Installing the Nebulizer and Nebulization Chamber Assembly

Connecting the Siphon Overflow

1 Attach the Tygon™ drain tube assembly to the end of the siphon tube using
2 Locate the tube in such a way that condensed solvent can flow freely from
3 Ensure that no siphon liquid leak could affect detector performances or
Toxic and hazardous solvents
WA R N I N G
The handling of solvents and reagents can hold health risks.
=> When opening capillary or tube fittings solvents may leak out.
=> Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
46
the tapered hose connector and lead the tube to a waste container.
the siphon and ensure that the end of the tube is not immersed in the
collected liquid. Make sure that the liquid container is appropriate for the
solvents used.
create a laboratory pollution.
If the solvent that you are using is not compatible with Tygon™ (e.g. THF),
use instead a piece of Teflon tubing or any material that you know is
compatible with your solvent.
Agilent 1260 Infinity ELSD User Manual