e. Carefully separate the Acquisition board from the combination assembly
chassis at the front corners of the Acquisition board.
5. Remove the TekConnect circuit board: See Figure 6- -42, page 6- -64. Remove
the five T-15 Torxdrive screws securing the TekConnect board to the
combination assembly.
6. Remove a TekConnect bucket: See Figure 6- -42, page 6- -64. Remove the
three T-15 Torxdrive screws securing the TekConnect bucket to the Front
Panel Combination assembly.
Removal and Installation Procedures
6- 61