Description, continued
Condensate Removal
Sample Gas Outlet
Temperature
Measurement and
Display
Sample Gas Outlet
Temperature
Monitoring
Set Point Adjustment
Fig. 9
Circuit Diagram
26
The condensate that accumulates comes out via the condensate nozzle on the
heat exchanger 5 (see Fig. 8).
The condensate is removed automatically by the peristaltic pump, which is
incorporated as an optional extra.
The measuring gas outlet temperature is measured using the Pt-100 temperature
sensor in the cooling block 7 of the sample gas cooler, and displayed in digital
form in °C on temperature controller 9 (see Fig. 8). The sample gas outlet tempera-
ture is set in the factory to +3 °C.
The temperature controller sends out a status signal if the temperature rises
above / falls below the respective limit values of +3 °C ± 3 °C as set up in the
factory. This signal is present at a floating change-over contact that is rated up to
250 V AC / 2 A.
The set point for the sample gas outlet temperature can be set at the temperature
controller:
Press key P; the set point is displayed. Use arrow keys to adjust the set point.
Press key P; the new set point is stored.
Bridge = 230V Transformer = 115V
Mains L
F1
X1/4
X2/2
L
2A T
275V
R2
X2/1
U
L
X4/1
X4/3
1
Option
Pt 100
T
3
X4/2
X4/4
5 +
X3/3
X3/1
6 -
K1
Status 1
8
X3/2
9
X3/6
N
X3/4
A2
K1
Status 2
X3/5
Mains N
X1/3
Mains PE
X1/2
X1/1
SCC-C Sample Gas Cooler Operator's Manual
X2/3
X5/O/5
T
R1
X6/4
A1
X6/3
X5/O/1
X6/2
X6/1
Fan
M2
M
MV1
K1
~
X5/M/5
X5/M/1
X5/U/1
X5/U/4
X5/O/3
M
~
M1
C
S
R
X5/O/2
X5/M/2
C1
4,7µF
X5/U/2
X5/M/3
X5/O/4
X5/O/7
X5/O/6
SR25
SR25
M4
M3
M
M
~
~
X5/M/4
X5/M/7
X5/M/6
42/23-55 EN Rev. 1
X5/U/3

Advertisement

Table of Contents
loading

Table of Contents