Power Supply - ABB SCC-C Operator's Manual

Sample gas cooler
Table of Contents

Advertisement

Power Supply

Power Supply
Sample Gas Inlet Conditions
Sample Gas
Inlet Conditions
Scope of Delivery
Scope of Delivery
42/23-55 EN Rev. 1
Input Voltage
Power Consumption
Starting Current
Sample Gas Pressure
Sample Gas Flow
Sample Gas Inlet Temperature
Sample Gas Inlet Dew Point
Quantity
Description
1
Sample gas cooler
2
Operator's manuals (in English and in German language)
for the model with glass heat exchanger:
GL coupling nuts with inset 6 x 4 x 1 threads for a hose or pipe
(2 per heat exchanger)
The fixing brackets are fitted in the factory.
SCC-C Sample Gas Cooler Operator's Manual
230 VAC or 115 VAC, –15 to +10 %, 50 to 60 Hz
max. 200 VA
2.5 A at 230 VAC
Heat
Sample Gas Pressure p
Exchanger
without
Material
Peristaltic Pump
Glass
50 to 200 kPa
(0.5 to 2.0 bar)
PVDF
50 to 250 kPa
(0.5 to 2.5 bar)
Stainless
0.05 to 1 MPa
Steel
(0.5 to 10 bar)
1 x 250 l/h or 1 x 125 l/h or 2 x 125 l/h,
assuming sample gas pressure p
(1 bar) and 25 °C
140 °C
70 °C, max. 60 °C for heat exchanger HE250
where sample gas flow > 200 l/h
abs
with
50 to 150 kPa
(0.5 to 1.5 bar)
50 to 150 kPa
(0.5 to 1.5 bar)
= 100 kPa
abs
7