Indications

This chapter describes the iSleep 20+'s indication functions, messages shown in
the display window.
Power Fail Indication
I
TEM
Definition
Action
Indication
Reset
Internal Function Error
I
TEM
Definition
Doc. 004270 En-Eu T-3b
D
ESCRIPTION
A power fail indication will be given when the
iSleep 20+ is disconnected from the mains power
source more than 5 seconds.
The iSleep 20+ will terminate treatment.
The indication is asserted with an audible tone.
In order to reset the indication press the Start/
Stop button.
D
ESCRIPTION
An internal function error will be given when the
iSleep 20+ has an internal function failure. The
following error codes exist:
35: High temperature. Let the iSleep 20+ cool
down and restart. The ambient temperature
might be too high.