The iSleep 20+ will exclude any mask-off time and ramp time when registering
the patient operating time.
If the mask is not fitted properly to the patient, the iSleep 20+ may incor-
rectly detect that the mask is taken off and may change the delivered
pressure to the reduced mask-off level. In this case, check the mask fit-
ting and adjust it if necessary.
Snooze
The snooze function allows the
patient to pause the treatment. This is
done by shortly pressing the Start/
Stop button. When the snooze is
started, the display will lighten up and
the iSleep 20+ delivers a low flow.
In order to start the treatment again, start to breath in the mask or press the
Start/Stop button shortly. The iSleep 20+ will then increase the pressure to the
set level during up to 10 seconds.
The iSleep 20+ will turn off automatically after 10 minutes of inactivity once
the snooze function has been activated.
Once the snooze is activated, and the mask has not been taken off within
10 seconds, the pressure will rise to the pressure set before the snooze function
was activated.
Humidifier (optional)
The humidifier function is adjustable to provide additional humidity to the
patient air.
Automatic Restart After Power Fail
The iSleep 20+ will restart after a power failure with a duration shorter than 5
seconds.
Wake Up (optional)
The wake up function will initiate an alarm at a set time when activated. Press
the Start/Stop button to turn off the alarm.
20 Functions and Parameters of the iSleep 20+
iSleep 20+ users manual
Doc. 004270 En-Eu T-3b