Page 1
Your purchase of this Digital Thermometer marks a step forward for you into the field of precision measurements. Although this Thermometer is a complex and delicate instrument, its ruggedness will allow many years. Of use if proper operating techniques are developed.
1. FEATURES * Can match any standard type K sensor. * Fitted with standard K probe socket. * LCD display allows clear read out even at bright places. * LCD display provides low power consumption. * LSI-circuit use provides high reliability and durability. * High accuracy and wide measurement range.
3. SPECIFICATIONS 3-1 General Specification Environment Conditions: Operating Temperature: 0℃ to 40℃≤80% RH, non-condensing Storage Temperature: -10℃ to 60℃ ≤70% RH, battery removed Operating principle: dual slop integration Sample Rate: 2 times/sec for digital data Display: 3 1/2 digits LCD Display with max. Reading 1999, Automatic indication of functions.
Page 6
The " " is displayed when the battery is under the proper operation range Impedance: 10MΩ Dimension: 125.5(L)×72(W)×27(H)mm Weight: 145g approx. (battery included) TP-01 Thermocouple Probe: The sensor supplied with the instrument is an ultra-fast response naked-bead thermocouple TP-01 suitable for many applications but with a maximum operating temperature of 250℃/482℉(300℃/572℉...
Page 7
3-2 Technical Specification (Note: The series of thermometer has two models, and they are differentiated by A and B. Their accuracy, resolutions and measuring ranges are different. You can find the measuring range from the meter’s panel.) Measurement Range: -50.0℃ to 150.0℃; -58℉ to 302℉ (A) ℃...
Page 8
Accuracy: ±(% of reading + number of digits) at 18 C to 28 C (64 F to 82 with relative humidity to 80%. 3-2-1 Celsius scale Range accuracy ±6℃ ± -50.0℃ to -40.0℃ -50℃ to -40℃ 6℃ ±3℃ ±5℃ -39.9℃ to -20.0℃ -40℃...
Page 9
3-2-3 Fahrenheit scale Range accuracy ±8℉ ±7℉ -58℉ to -4℉ -58℉ to -4℉ ±4℉ ±4℉ -3℉ to 32℉ -4℉ to 32℉ ±2℉ ℉ ±1%±5 33℉ to 212℉ 32℉ to 932℉ ±4℉ ±0.75%±4℉ 213℉ to 302℉ 932℉ to 1382℉ >>>7<<<...
4. MEASURING PROCEDURE 4-1 Temperature Measurement ● For measurement of high temperatures, surfaces, semi-solids, liquids etc., a range of hand-held probes is available (such as TP-02A…) or, it required, any suitable probe of the K type can be used. ● Insert the sensor plug into the socket at top edge of the instrument, taking care to observe the correct polarity.
Page 11
4-2 Consideration When the sensor is first plugged into the thermometer, or if the sensor is changed, the plug must be allowed to stabilize at the temperature of the socket, which is in thermal contact with the cold junction compensation device, for greatest accuracy is to be achieved.
Page 12
IN 43710 Measurement Temp. Allowable Error ±3% 0℃ TO 400℃ ±0.75% 400 TO 1100℃ JIS C1602-1981 Measurement Class Allowable Error Temperature ±1.5℃ or ±0.4% of Measured Temp. >0℃<1000℃ ±2.5℃ or ±0.75% of Measured Temp. >0℃<1200℃ 0.75 ±2.5℃ or ±1.5% of Measured Temp. >-200℃<0℃...
5. MAINTENANCE 5-1 General maintenance To keep the instrument clean, wipe the case with a dry cloth and detergent, do not use abrasives or solvents. Any adjustment maintenance and repair shall be by a skilled person about temperature technology. To maintain a thermocouple in good condition, shall observe the following items.
symbol will appear on the LCD display and the battery need to be replaced. Slide the battery cover away from the instrument and remove the battery. Replace with 9V battery and reinstall the cover 6. OPTIONAL PROBE (TYPE K) TP-02A (optional): ℃...