Table of Contents

Advertisement

Quick Links

Measurement Guide
3GPP Signal Analyzer
for Anritsu RF and Microwave Handheld Instruments
BTS Master™, Cell Master™, Spectrum Master™, LMR Master™
GSM/GPRS/EDGE
W-CDMA/HSPA+
TD-SCDMA/HSPA+
LTE/LTE-A
TD-LTE/LTE-A
For some models, RF, Demod, and OTA are combined as a single option.
GSM/GPRS/EDGE
W-CDMA/HSPA+
TD-SCDMA/HSPA+
LTE/LTE-A (FDD and TDD)
Not all instrument models offer every option or every measurement within a given
Note
option. Please refer to the Technical Data Sheet of your instrument for available
options and measurements within the options.
Anritsu Company
490 Jarvis Drive
Morgan Hill, CA 95037-2809
USA
http://www.anritsu.com
RF
Option 40
Option 44
Option 60
Option 541
Option 551
Demod
Option 41
Option 65
Option 61
Option 542
Option 552
Option 880
Option 881
Option 882
Option 883
Part Number: 10580-00234
Published: September 2017
Copyright 2017 Anritsu Company
OTA
N/A
Option 35
Option 38
Option 546
Option 556
Revision: Y

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3GPP and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Anritsu 3GPP

  • Page 1 Measurement Guide 3GPP Signal Analyzer for Anritsu RF and Microwave Handheld Instruments BTS Master™, Cell Master™, Spectrum Master™, LMR Master™ Demod GSM/GPRS/EDGE Option 40 Option 41 W-CDMA/HSPA+ Option 44 Option 65 Option 35 TD-SCDMA/HSPA+ Option 60 Option 61 Option 38...
  • Page 2 BTS Master, Cell Master, LMR Master, and Spectrum Master are trademarks of Anritsu Company. Notice Anritsu Company has prepared this manual for use by Anritsu Company personnel and customers as a guide for the proper installation, operation and maintenance of Anritsu Company equipment and computer programs.
  • Page 3: Table Of Contents

    Introduction ..........1-1 3GPP Signal Analysis Overview....... . . 1-1 GSM/GPRS/EDGE .
  • Page 4 3-10 Amplitude Menu ..........3-27 Contents-2 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 5 Measurement Setup ........4-12 3GPP MG PN: 10580-00234 Rev.
  • Page 6 Modulation Summary ........5-10 Modulation Measurement Setups ......5-11 Contents-4 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 7 5-18 Limit Menu ..........5-47 3GPP MG PN: 10580-00234 Rev.
  • Page 8 Trigger Menu ..........6-33 Contents-6 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 9 Anritsu easyMap Tools ........
  • Page 10 3GPP Messages ........
  • Page 11: Introduction

    Chapter 1 — General Information Introduction This Measurement Guide documents 3GPP signal analysis for the following Anritsu instruments: • BTS Master • Cell Master • Spectrum Master • LMR Master Not all instrument models offer every option. Please refer to the Technical Data...
  • Page 12: Td-Scdma/Hspa+

    1-2 3GPP Signal Analysis Overview General Information TD-SCDMA/HSPA+ TD-SCDMA/HSPA+ signal analysis is described in Chapter 4, “TD-SCDMA/HSPA+ Signal Analyzer”. • Option 882: TD-SCDMA/HSPA+ Measurements (requires Option 9) • Option 60*: TD-SCDMA/HSPA+ RF Measurements • Option 61*: TD-SCDMA/HSPA+ Demodulation • Option 38*: TD-SCDMA/HSPA+ OTA Measurements LTE signal analysis is described in Chapter 5, “LTE Signal...
  • Page 13: Selecting A Measurement Mode

    Selector list box. Highlight the desired measurement mode using the Up or Down arrow keys or the rotary knob and press Enter. The list of available applications depends on the options that are installed and activated on your instrument. Figure 1-1. Mode Selector List Box 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 14: Contacting Anritsu

    Note this measurement guide, depending on instrument model, firmware version, and installed options. Contacting Anritsu To contact Anritsu, visit the following URL and select the services in your region: http://www.anritsu.com/contact-us. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 15: Introduction

    This chapter describes the menus in GSM/GPRS/EDGE Signal Analyzer mode. Screen capture images are provided as examples. The image and measurement Note details shown on your instrument may differ from the examples in this measurement guide. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 16: Measurement Setup

    Press the Measurements main menu key to select measurement display options. To display Spectrum, press the Spectrum submenu key. Press the Channel Spectrum submenu key for a single channel (Figure 2-1), or the Multi-Channel Spectrum submenu key to display the multi-channel spectrum (Figure 2-2). PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 17 GSM/GPRS/EDGE Signal Analyzer 2-4 Measurement Display Figure 2-1. GSM Single Channel Measurement Figure 2-2. GSM Multi-Channel Measurement 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 18 To display Power versus Time (Slot) press the Power versus Time (Slot) submenu key to activate the Power versus Time (Slot) measurement (Figure 2-4). The mask is according to the 3GPP TS 05.05 specification. The first slot information is displayed. Figure 2-4. GSM Power vs. Time (Slot) Measurement PN: 10580-00234 Rev. Y...
  • Page 19: Demodulator

    6. Press the Demodulator submenu key. Figure 2-5. GSM Demodulator Measurement Using multi-channel spectrum, channel cursor, select the channel and press the Note Demodulator submenu key and the unit will demodulate the selected channel. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 20 2-5 Demodulator GSM/GPRS/EDGE Signal Analyzer Figure 2-6. EDGE Demodulator Measurement To display the GSM/EDGE Summary screen (Figure 2-7), press the GSM/EDGE Summary soft key. Figure 2-7. GSM/EDGE Summary PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 21: Gsm/Gprs/Edge Mode Pass/Fail

    5. Use the rotary knob or Up/Down arrow keys to select the applicable test set and to activate the measurement. Refer to the Master Software Tools documentation for information on creating a custom pass/fail test set. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 22: Measurement Results

    This measurement is performed for 8PSK modulated signals (EDGE) only. Origin Offset (dB) Origin Offset is the carrier leakage component of the measured signal in dB and this measurement is applicable to EDGE signal only. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 23 The modulation type can be GMSK (for GSM signals) or 8PSK (for EDGE signals). Mag Err (rms) The RMS of the magnitude error between the received signal and an ideal reconstructed reference signal of the first active slot in %. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 24: Gsm/Gprs/Edge Menus

    Pass/Fail Mode Sweep I/Q View Select Sweep Pass/Fail Vector Const Test Continuous Single Trigger Reset Sweep ScriptMaster Back ScriptMaster Start Test Change File Header Test_Seq_1 Back Skip Exit Figure 2-9. GSM/GPRS/EDGE Menu Layout 2-10 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 25: Freq (Frequency) Menu

    Set CF To Closest Channel: Changes the center frequency to the closest Increment channel. Channel Decrement Channel: Decreases the channel number one channel. Increment Channel: Increases the channel number one channel. Figure 2-10. GSM/GPRS/EDGE Freq Menu 3GPP MG PN: 10580-00234 Rev. Y 2-11...
  • Page 26: Amplitude Menu

    GSM or EDGE signal automatically. Selecting GSM or EDGE sets the instrument to measure only a GSM or EDGE signal. GSM/EDGE Select Auto EDGE Figure 2-12. GSM/GPRS/EDGE Setup Menu 2-12 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 27: Measurements Menu

    Power vs. Power versus Time (Slot): Displays the first active slot of the GSM/EDGE Time signal capture. The mask is as specified in 3GPP TS 05.05. The screen also (Frame) displays Channel Power, Occupied Bandwidth, Frequency Error in PPM and Power vs.
  • Page 28: Spectrum Menu

    Press the Enter key to select. Back Back: Returns to the Spectrum Menu. Back: Returns to the “Measurements Menu”. MC Spectrum Channel Cursor Freq Cursor 1.000 GHz Back Figure 2-14. Spectrum Menu 2-14 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 29: Demodulator Menu

    2-12 Measurements Menu Demodulator Menu Key Sequence: Measurements > Demodulator I/Q View: Allows selection between Vector or Const. Demodulator Back: Returns to the “Measurements Menu”. I/Q View Vector Const Back Figure 2-15. GSM/GPRS/EDGE Freq Menu 3GPP MG PN: 10580-00234 Rev. Y 2-15...
  • Page 30: Pass/Fail Mode Menu

    Exit: Ends the ScriptMaster measurement process. Returns the instrument to the previous measurement setup. Back Back: Returns to the “Measurements Menu”. ScriptMaster Start Test Change File Header Test_Seq_1 Skip Exit Figure 2-16. GSM/GPRS/EDGE Pass/Fail Menu 2-16 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 31: Marker Menu

    This menu is not available in GMS/GPRS/EDGE measurement mode. 2-17 Limit Menu This menu is not available in GMS/GPRS/EDGE measurement mode. 2-18 Other Menus Preset, File, Mode and System are described in the instrument User Guide. 3GPP MG PN: 10580-00234 Rev. Y 2-17...
  • Page 32 2-18 Other Menus GSM/GPRS/EDGE Signal Analyzer 2-18 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 33: Introduction

    B equipment to the RF In connector of the unit using a coupler or attenuator. The maximum input damage level of the RF In port is +30 dBm. To prevent Note damage, always use a coupler or high power attenuator. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 34: W-Cdma/Hspa+ Measurements

    Carrier Feedthrough Carrier Feedthrough measures the amount of unmodulated signal that is leaking through the transmitter and is displayed in the Code Domain Power display. The W-CDMA 3GPP specification does not specify carrier feedthrough measurement. Code Domain Power displays how much of the channel power is in each Orthogonal Variable Spreading Factor (OVSF code).
  • Page 35 Spreading Factor (OVSF) According to the 3GPP standard the spreading factor can be from 4 to 512, and the instrument can be set to a maximum spreading factor of 256 or 512. Freq Error Frequency error is the difference between the received center frequency and the specified center frequency.
  • Page 36 PCDE is the maximum value for the code domain error for all codes (both active and inactive). In the 3GPP standard, in order to address the possibility of uneven error power distribution in W-CDMA, the EVM measurement has been supplemented with PCDE. The 3GPP standard requires that the PCDE not exceed –33 dB at a spreading factor of 256.
  • Page 37: General Measurement Setups

    In a W-CDMA system, the number of chips per data symbol is called the Spreading Factor. The lower the spreading factor the higher the data rate. According to the 3GPP standard, the spreading factor can vary from 4 to 512 and the maximum spreading factor is either 256 or 512.
  • Page 38: Threshold Setup

    The ACLR measurement uses the filtered channel power to determine the ACLR values and it is listed as filtered on the display. In all other screens the unfiltered channel power is displayed as channel power. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 39: W-Cdma/Hspa+ Rf Measurements

    1. Set the measurement frequency by using one of the methods listed in the User Guide. 2. Press the Measurements main menu key and the RF Measurements submenu key. 3. Press the Band Spectrum submenu key to display the band spectrum (Figure 3-1). 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 40: Channel Spectrum Setup

    5. Press the Channel Spectrum submenu key to activate the spectrum measurement (Figure 3-2). Using the Band Spectrum cursor, select the desired channel and the unit will Note automatically display the measurements for the selected channel when the Channel Spectrum key is selected. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 41: Aclr Measurement Setup

    The 3GPP standard specifies one main channel and two adjacent channels. The ACLR screen displays the main channel power and the power of two adjacent channels on each side as a bar graph.
  • Page 42 Using the Band Spectrum cursor, select the required channel and press the ACLR Note submenu key. The measurement will be displayed. 5. Press the ACLR submenu key again and select one main channel and two adjacent channels (Figure 3-3). Figure 3-3. ACLR Measurement 3-10 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 43: Aclr Multi-Channel Aclr Procedure

    The Spectral Emission Mask displays the selected signal and the mask as defined in the 3GPP specification. The mask varies depending upon the input signal. The instrument also indicates if the signal is within the specified limits by displaying PASSED or FAILED. The emission mask is also displayed in a table format with different frequency ranges and whether the signal PASSED/FAILED in that region.
  • Page 44: Spectral Emission Mask Procedure

    2. Press the Measurements main menu key. 3. Press the RF Measurements submenu key. 4. Press the Spectral Emission Mask submenu key to activate the Spectral Emission Mask measurement (Figure 3-5). Figure 3-5. Spectral Emission Mask Measurement 3-12 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 45: Rf Summary

    W-CDMA/HSPA+ signal. The parameters displayed in the RF summary table are Channel Power, Carrier Frequency, Frequency Error, Spectral emission Pass/Fail criteria, Occupied Bandwidth, Peak to Average Power, ACLR at –10 MHz, –5 MHz, 5 MHz and 10 MHz channels. 3GPP MG PN: 10580-00234 Rev. Y 3-13...
  • Page 46: Demodulator

    OVSF codes. The demodulator also displays CPICH, P-CCPCH, S-CCPCH, PICH, P-SCH and S-SCH power in the table format. For W-CDMA/HSPA+ Demodulator, the HSPA+ codes are also displayed. 3-14 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 47: Code Domain Power (Cdp) Procedure

    16. Press the Back submenu key to go back to the CDP measurement. The blue color block on the CDP screen represents the selected zoom codes and Note the same codes are displayed in the zoom screen. 3GPP MG PN: 10580-00234 Rev. Y 3-15...
  • Page 48 CDP measurements table. Markers can be used to read the individual code power, symbol EVM (@ EVM) Note and type of code and can be activated in all the W-CDMA/HSPA+ measurements. 3-16 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 49: Hspa+ Setup

    The maximum time is 72 hours. 5. Use the cursor to select the desired code. The code parameters are displayed on the screen. 3GPP MG PN: 10580-00234 Rev. Y 3-17...
  • Page 50: Codogram Setup

    7. Press the Threshold submenu key to manually set the Threshold level which determines which codes are active. The default value is –30dB. 8. Press the Measurements main menu key. 9. Press the Demodulator submenu key to list the Demod menu. 3-18 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 51 The blue color block on the Codogram screen represents the selected zoom codes Note and the same codes are displayed in the zoom screen. Save the data before making any measurements, otherwise the data will be lost. Figure 3-10. Codogram Measurement 3GPP MG PN: 10580-00234 Rev. Y 3-19...
  • Page 52: Over-The-Air Measurements

    On/Off submenu key to turn the selected code on or off. The Code Lock submenu key locks the code, so that the code will not change with each update. Refer to “Over-The-Air Menu” on page 3-38 for a description of OTA measurement submenus. 3-20 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 53 W-CDMA/HSPA+ Signal Analyzer 3-6 Over-the-Air Measurements Figure 3-11. OTA Measurement Summary 3GPP MG PN: 10580-00234 Rev. Y 3-21...
  • Page 54: W-Cdma Summary Setup

    16 will be displayed in all the views. Set the PICH code to show accurate results. 9. Press the Threshold submenu key to manually set the Threshold level to determine which codes are active. The default value is –30dB. 3-22 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 55: Pass/Fail Mode Setup

    11. Press the W-CDMA Summary submenu key. Pass/Fail Mode Setup The instrument stores the five test models specified in the 3GPP specification (TS 125.141) for testing base station performance and recalls these models for quick easy measurements. After selection of a test model, the instrument displays test results in tabular format with clear PASS or FAIL indications that include min/max threshold.
  • Page 56: W-Cdma/Hspa+ Menus

    Select Next # of Freq Err Avg Cnt Adjacent Channels Band Summary Channel Width Sync Type 5.000 MHz CPICH No CPICH Back Back Back Back Figure 3-14. W-CDMA/HSPA+ Menu Layout (1 of 2) 3-24 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 57 Rel Pwr CPICH EC/Io EMF Units Horiz Display Sort By Chips uSec Code Power Exit Limits Distance Code Lock Reset 6.00 V/m Back Back Back Figure 3-15. W-CDMA/HSPA+ Menu Layout (2 of 2) 3GPP MG PN: 10580-00234 Rev. Y 3-25...
  • Page 58: Freq (Frequency) Menu

    Set CF To Closest Channel: Changes the center frequency to the closest channel. Increment Decrement Channel: Decreases the channel number by one. Channel Increment Channel: Increases the channel number by one. Figure 3-16. W-CDMA/HSPA+ Freq Menu 3-26 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 59: Amplitude Menu

    Adjust Range: Press this submenu key to perform a single reference level adjustment. Auto Range is automatically turned Off. Unit: Changes the unit of measure for the Y-axis between dBm and W. Units External Loss External Gain Backspace Figure 3-17. W-CDMA/HSPA+ Amplitude Menu 3GPP MG PN: 10580-00234 Rev. Y 3-27...
  • Page 60: Setup Menu

    PICH Code Indicator Channel and enter the desired code. The default value is 16. More: Lists the “Setup 2/2 Menu” on page 3-29 to continue measurement setup. More Figure 3-18. W-CDMA/HSPA+ Setup Menu (1/2) 3-28 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 61: Setup 2/2 Menu

    Sync Type: Selects between CPICH or No CPICH for synchronization. Freq Err Avg Freq Err Avg Cnt Back: Returns to the “Setup 1/2 Menu” on page 3-28. Sync Type CPICH No CPICH Back Figure 3-19. W-CDMA/HSPA+ Setup Menu (2/2) 3GPP MG PN: 10580-00234 Rev. Y 3-29...
  • Page 62: Measurements Menu

    “Pass/Fail Mode Menu” on page 3-41. The instrument saves the five test model conditions specified in the 3GPP specification to test the base station. After the selected test model, the unit displays whether the base station passed or failed the test. Using Master...
  • Page 63: Rf Measurement Menu

    Mask RF Summary: Displays the RF measurements in table form. Spectral Emission Back: Returns to the “Measurements Menu” on page 3-30. Summary Summary Back Figure 3-21. W-CDMA/HSPA+ RF Measurement Menu 3GPP MG PN: 10580-00234 Rev. Y 3-31...
  • Page 64: Band Spectrum Menu

    Span: Select either a 5 MHz or 10 MHz span. Channel Spectrum Back: Returns to the “RF Measurement Menu” on page 3-31. Span 5 MHz 10 MHz Back Figure 3-23. W-CDMA/HSPA+ Channel Spectrum Menu 3-32 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 65: Aclr Menu

    Channel Width: Set the bandwidth for the channels for the number of # of Main Channels adjacent channels selected. Back: Returns to the “RF Measurement Menu” on page 3-31. Select # of Adjacent Channels Channel Width 5.000 MHz Back Figure 3-24. W-CDMA/HSPA+ ACLR Menu 3GPP MG PN: 10580-00234 Rev. Y 3-33...
  • Page 66: Demodulator Menu

    OVSF codes and the bottom one displays the selected OVSF zoom codes. Modulation Summary: Displays the demodulation parameters in the table format. Back: Returns to the “Measurements Menu” on page 3-30. Figure 3-25. W-CDMA/HSPA+ Demodulator Menu 3-34 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 67: Cdp Menu

    2, enter 2. Control Channels: Select a control channel mode between Rel, Abs, and Delta. Zoom Start Back: Returns to the “Demodulator Menu” on page 3-34. Control Channels Delta Back Figure 3-26. W-CDMA/HSPA+ CDP Menu 3GPP MG PN: 10580-00234 Rev. Y 3-35...
  • Page 68: Hsdpa Menu

    Single Sweep Time Control Channels: Select a control channel mode between Rel, Abs, and Delta. IQ Persistence Back: Returns to the “Demodulator Menu” on page 3-34. Control Channels Delta Back Figure 3-27. HSPA+ Menu 3-36 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 69: Codogram Menu

    Total Time Back: Returns to the “Demodulator Menu” on page 3-34. 240 s Single Sweep Time Back Figure 3-28. W-CDMA/HSPA+ Codogram Menu 3GPP MG PN: 10580-00234 Rev. Y 3-37...
  • Page 70: Over-The-Air Menu

    EMF Measurement mode. Once the EMF Measurement mode is active, this key opens the “W-CDMA EMF Menu” on page 8-10. Back: Returns to the “Measurements Menu” on page 3-30. Multipath Back Figure 3-29. W-CDMA/HSPA+ Over-The-Air Menu 3-38 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 71: Ota Setup Menu

    Code Lock Reset: Reset the measurement screen. Back: Returns to the “Over-The-Air Menu” on page 3-38. Code Lock Display Unit CPICH EC/Io Sort By Code Power Code Lock Reset Back Figure 3-30. W-CDMA/HSPA+ Over-The-Air (OTA) Setup Menu 3GPP MG PN: 10580-00234 Rev. Y 3-39...
  • Page 72: Multipath Menu

    Vert Display RSCP Rel Pwr Horiz Display Chips uSec Distance Back Figure 3-31. W-CDMA/HSPA+ Multipath Menu EMF Menu Key Sequence: Measurements > OTA > EMF Refer to “W-CDMA EMF Menu” on page 8-10. 3-40 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 73: Pass/Fail Mode Menu

    Text Entry menu to name the saved file after the ScriptMaster Test_Seq_1 measurement process is complete. Skip: Skips over a ScripMaster instruction. Skip Exit: Ends the ScriptMaster measurement process. Returns the instrument to the previous measurement setup. Exit Figure 3-33. W-CDMA/HSPA+ ScriptMaster Menu 3GPP MG PN: 10580-00234 Rev. Y 3-41...
  • Page 74: Marker Menu

    Trigger Sweep: Pressing this submenu key causes the instrument to make a single sweep when the instrument is in single sweep mode. This key has Trigger no function when the instrument is in continuous sweep mode. Sweep Figure 3-35. W-CDMA/HSPA+ Sweep Menu 3-42 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 75: Measure Menu

    This menu is not available in W-CDMA/HSPA+ measurement mode. 3-17 Limit Menu This menu is not available in W-CDMA/HSPA+ measurement mode. 3-18 Other Menus Preset, File, Mode and System are described in the instrument User Guide. 3GPP MG PN: 10580-00234 Rev. Y 3-43...
  • Page 76 3-18 Other Menus W-CDMA/HSPA+ Signal Analyzer 3-44 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 77: Introduction

    Span is 5 MHz when Number of Carrier is set to 1. For all other values, span is set to 1.6 MHz. DwPTS Pwr The power in the Downlink Pilot slot, excluding the gap. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 78: Measurement Setup

    This value is N/A when Number of Carriers is set to 1. Measurement Setup Refer to the User Guide for general instrument setup instructions, and continue with the following setups for the specific RF measurements. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 79: Channel Spectrum

    4. Press the RF Measurements submenu key. 5. Press the Channel Spectrum submenu key to activate this RF measurement view (see Figure 4-1). 6. Press the Back submenu key again to return to the Measurements menu. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 80: Power Versus Time

    (0 to 6), then press the Enter key. c. Press the Back submenu key to return to the RF Measurements menu. d. Press the Back submenu key again to return to the Measurements menu. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 81 TD-SCDMA/HSPA+ Signal Analyzer 4-3 TD-SCDMA/HSPA+ RF Measurements Figure 4-2. TD-SCDMA/HSPA+ Power vs. Time: Sub-Frame Measurement Figure 4-3. TD-SCDMA/HSPA+ Power vs. Time: Slot Measurement 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 82: Rf Summary

    2. Press the Measurements main menu key. 3. Press the RF Measurements submenu key. 4. In the RF Measurements menu, press the RF Summary submenu key to view the primary RF measurements in a table. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 83: Td-Scdma/Hspa+ Demodulator

    GPS reference. Tau is the timing delay of the frame starting point in reference to the occurrence of a trigger. When no trigger is used, Tau values are relative to the most dominant SYNC-DL code. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 84: Measurement Setup

    Press the Max Users submenu key to open the Maximum Users selection window and menu. Use the rotary knob or press the touch screen to highlight Auto or the desired value (2 to 16), then press Enter. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 85 Code Domain Power (CDP) and Code Domain Error (CDE) are displayed with color coding as described in Table 4-1: Table 4-1. Color Legend for CDP and CDE Display Color Code Type Orange Active Codes Gray Idle Codes Light Blue Code Domain Errors 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 86: Td-Scdma/Hspa+ Over-The-Air Measurements

    The power in the Downlink Pilot slot, excluding the gap. Pilot Dominance Pilot dominance is a measure of the strength of the strongest code compared to the next strongest code in the same channel. 4-10 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 87: Measurement Setup

    3. Press the Code Scan submenu key to activate this measurement view. Tau Scan 1. Press the Measurements main menu key. 2. Press the OTA submenu key. 3. Press the Tau Scan submenu key to activate this measurement view. 3GPP MG PN: 10580-00234 Rev. Y 4-11...
  • Page 88: Td-Scdma/Hspa+ Pass/Fail Measurements

    5. Press the Reset submenu key to begin a new pass/fail test measurement (see Figure 4-7). 6. Press the Back submenu key to return to the Measurements menu. Figure 4-7. TD-SCDMA/HSPA+ Pass/Fail Measurements 4-12 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 89: Td-Scdma/Hspa+ Menus

    Scrambling Code Max Users Trigger Type Auto Auto Auto Auto No Trig Ext Trigger Polarity Rising Falling Tau Offset Cancel Cancel Cancel Cancel 0 ms Back Figure 4-8. TD-SCDMA/HSPA+ Menu Layout (1 of 2) 3GPP MG PN: 10580-00234 Rev. Y 4-13...
  • Page 90 Code Scan Pass/Fail CDP Data Test CDP Units Relative Absolute Tau Scan Reset Table CDP Marker Modulation Record Back Summary Back Back Run/Hold Hold Back Figure 4-9. TD-SCDMA/HSPA+ Menu Layout (2 of 2) 4-14 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 91: Freq (Frequency) Menu

    Channel Decrement Channel: Decrements the channel by one channel. Decrement Increment Channel: Increments the channel by one channel. Channel Increment Channel Figure 4-10. TD-SCDMA/HSPA+ Freq Menu 3GPP MG PN: 10580-00234 Rev. Y 4-15...
  • Page 92: Amplitude Menu

    Adjust Range: Press this submenu key to perform a single reference level adjustment. Auto Range is automatically turned Off. Units Units: Press this submenu key to view the power measurements in dBm or watts. External Loss External Gain Backspace Figure 4-11. TD-SCDMA/HSPA+ Amplitude Menu 4-16 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 93: Setup Menu

    Meas Speed: Press this submenu key to select the desired scan speed: Fast, Normal, or Slow. The active state is underlined on the face of the submenu key. More: Opens the “Advanced Settings Menu” on page 4-20. Figure 4-12. TD-SCDMA/HSPA+ Setup Menu 3GPP MG PN: 10580-00234 Rev. Y 4-17...
  • Page 94: Slot Selection Menu

    Tau value. Rising Falling Back: Press this submenu key to return to “Setup Menu” on page 4-17. Tau Offset 0 ms Back Figure 4-14. TD-SCDMA/HSPA+ Trigger Menu 4-18 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 95: Sync-Dl Code Menu

    The Auto submenu key provides a one-touch method of Auto selecting the Auto setting. Cancel: Disregards any entry and returns to “Setup Menu” on page 4-17. Cancel Figure 4-17. TD-SCDMA/HSPA+ Max Users Menu 3GPP MG PN: 10580-00234 Rev. Y 4-19...
  • Page 96: Advanced Settings Menu

    Tau Offset that may be entered. 0 ms Back: Press this submenu key to return to “Setup Menu” on page 4-17. Back Figure 4-18. TD-SCDMA/HSPA+ Advanced Settings Menu Figure 4-19. Modulation Type Window 4-20 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 97: Measurements Menu

    Save Measurement: Press this submenu key to save a measurement. You may accept the default filename or enter your own filename. Refer to your instrument’s User Guide for information on file management functions. Figure 4-20. TD-SCDMA/HSPA+ Measurements Menu 3GPP MG PN: 10580-00234 Rev. Y 4-21...
  • Page 98: Rf Measurements Menu

    Left Channel Power Summary Right Channel Power Left Channel Occ BW Right Channel Occ BW Back Back: Press this submenu key to return to “Measurements Menu” on page 4-21. Figure 4-21. TD-SCDMA/HSPA+ RF Measurement Menu 4-22 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 99: Power Vs. Time Menu

    Mask: Press this submenu key to turn the mask On and Off. Back Back: Press this submenu key to return to the “Measurements Menu” on page 4-21. Slot Selection Auto Cancel Figure 4-22. TD-SCDMA/HSPA+ Power vs. Time Menu 3GPP MG PN: 10580-00234 Rev. Y 4-23...
  • Page 100: Demodulator Menu

    Noise Floor CDP Units Carrier Feed Through Relative Absolute Peak CDE CDP Marker Back: Press this submenu key to return to the “Measurements Menu” on page 4-21. Back Figure 4-23. TD-SCDMA/HSPA+ Demodulator Menu 4-24 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 101: Over-The-Air Menu

    Back: Press this submenu key to return to the “Measurements Menu” on page 4-21. Back Figure 4-24. TD-SCDMA/HSPA+ Over the Air Menu 3GPP MG PN: 10580-00234 Rev. Y 4-25...
  • Page 102: Pass Fail Mode Menu

    Reset: Press this submenu key to restart the measurement or to begin a Reset new pass/fail test measurement. Back: Press this submenu key to return to the “Measurements Menu” on page 4-21. Back Figure 4-25. TD-SCDMA/HSPA+ Pass Fail Mode Menu 4-26 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 103: Sweep Menu

    This menu is not available in TD-SCDMA/HSPA+ measurement mode. 4-15 Limit Menu This menu is not available in TD-SCDMA/HSPA+ measurement mode. 4-16 Other Menus Preset, File, Mode and System are described in the User Guide. 3GPP MG PN: 10580-00234 Rev. Y 4-27...
  • Page 104 4-16 Other Menus TD-SCDMA/HSPA+ Signal Analyzer 4-28 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 105: Introduction

    The instrument also automatically selects PBCH only when OTA measurements Note are selected. PBCH only is also necessary when connecting directly to the transmitter and MIMO is used, when measuring live traffic and assuming that precoding is used in that eNodeB. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 106: Antenna Status

    Constellation, Control Channel Power, Modulation Summary and LTE Summary. The antennas icons are displayed at the lower-left corner of the display screen (Figure 5-1). No Signal Single Antenna (Tx1, Tx2, Tx3, or Tx4) Two to Four Antennas (Tx Diversity) Figure 5-1. Antenna Status PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 107: Lte Rf Measurements

    The Spectral Emission Mask (SEM) measurement supports the testing for “Operating Band Unwanted Emissions” described in the 3GPP TS 36.141 Base Station Conformance testing document. There is support for Category A and Category B (Option 1 only) masks which are automatically selected based on the current carrier frequency/channel and BW values.
  • Page 108: Rf Measurement Setups

    4. Press the Channel Spectrum submenu key again to select the Channel Spectrum menu, where you can adjust the span. Figure 5-2. Channel Spectrum 5. Press the Back submenu key to return to the RF menu. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 109 2. Press the RF submenu key. 3. Press the ACLR submenu key to activate ACLR display and measurement (Figure 5-3). Figure 5-3. Adjacent Channel Leakage Ratio 4. Press the Back submenu key to return to the Measurements menu. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 110 5. To view the measurement data in a table format, press the Summary Table submenu key so On is underlined. Figure 5-4. Spectral Emission Mask 6. Press the Back submenu key to return to the RF menu. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 111 3. Press the RF Summary submenu key to activate and display the RF measurements in table format (Figure 5-5). Figure 5-5. RF Summary 4. Press the Back submenu key to return to the Measurements menu. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 112: Lte Modulation Measurements

    Option 886 is installed and activated on your instrument (see Figure 5-6 on page 5-11). Cell ID Cell identifying information sent by the transmitter in the sync signal. Measurement Progress Bar This progress bar indicates the sub-frame that is currently being measured. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 113: On Page

    Reset and the second number is the current measured value. Freq Error (ppm) Freq Error displayed in parts per million (ppm). Cell ID Cell identifying information sent by the transmitter in the sync signal. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 114: Control Channel Power

    Modulation Summary displays the Ref Signal (RS) Power, Sync Signal (SS) Power, EVM (rms), Freq Error (Hz and ppm), Cell ID and PBCH Power in a table format. Refer to individual modulation measurement descriptions for additional detail on each measurement. 5-10 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 115: Modulation Measurement Setups

    This is useful in the monitoring of changes in traffic loading on the Note base station. To measure the power levels of the physical layer Control Channels, use the “Control Channel Power” measurement. 3GPP MG PN: 10580-00234 Rev. Y 5-11...
  • Page 116 Constellation Measurement View Support for 256-QAM modulation as shown in the data legend above is available Note only with Option 886. 5. Press the Back submenu key to return to the Modulation menu. 5-12 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 117 5-8). Press the key again to view or change the Display Mode and Control Channel EVM settings. Figure 5-8. Control Channel Power 4. Press the Back submenu key to return to the Modulation menu. 3GPP MG PN: 10580-00234 Rev. Y 5-13...
  • Page 118 4x4 MIMO, the display may look similar to Figure 5-10. 4. Press the Back submenu key to return to the Measurements menu. Figure 5-9. Tx Time Alignment (4x4 MIMO Configuration) Figure 5-10. Tx Time Alignment 5-14 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 119: Measurements In Table Format

    3. Press the Modulation Summary submenu key to activate and display the modulation measurements in table format (Figure 5-11). Figure 5-11. Modulation Summary 4. Press the Back submenu key to return to the Measurements menu. 3GPP MG PN: 10580-00234 Rev. Y 5-15...
  • Page 120: Lte Over-The-Air (Ota) Measurements

    The PBCH Modulation Results table is displayed when the Show Mod Results setting is set to On. Refer to “OTA Scanner Menu” on page 5-39. The displayed values are the same as described under “Constellation” on page 5-9. 5-16 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 121: Mapping (Coverage Mapping)

    TX antennas to two RX antennas on the same frequency and time, but different reference signals. Your Anritsu test instrument supports 2x2 and 4x4 MIMO configurations. Note that 4x4 MIMO may not be available if your instrument is loaded with older firmware. Refer to your instrument User Guide for instructions on updating firmware.
  • Page 122: Emf (Option 444 Only)

    This is the frequency error displayed in parts per million. Time Alignment Error (TAE) is a measurement of the differential transmission delay between the Component Carriers. The 3GPP specification identifies the limits of the delays for the system to operate properly.
  • Page 123 5-20. Press the Tx Test key again to display the OTA Tx Test menu. 4. Press the Show Mod Results key to display or hide the PBCH Modulation Results (Strongest SS) table. 3GPP MG PN: 10580-00234 Rev. Y 5-19...
  • Page 124 Some instruments require the LTE demod option to toggle this submenu to On. Figure 5-13. Over-the-Air Tx Test Measurement Mapping Refer to Chapter 7, “LTE and TD-LTE Coverage Mapping”. 5-20 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 125 LTE Signal Analyzer 5-5 LTE Over-the-Air (OTA) Measurements Carrier Aggregation Your Anritsu test instrument supports single-input single-output (SISO) as well as 2x2 and 4x4 multiple-input multiple-output (MIMO) systems. If a single antenna is active, the Component Carrier measurements illustrated in...
  • Page 126: Pass/Fail Tests

    Software Tools (MST) User Guide for creating new Pass/Fail tests. b. Press the Reset submenu key to erase a previous measurement/s and restart the Pass/Fail test. Figure 5-15. Past/Fail Test 3. Press the Back submenu key to return to the Measurements menu. 5-22 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 127: Lte Summary

    Physical Control Format Indicator Channel Power. This channel provides information to enable the UE to decode the PDSCH (Physical Downlink Shared Channel). Spectral Emission Mask Displays a Pass/Fail status for the signal measured against the selected mask. 3GPP MG PN: 10580-00234 Rev. Y 5-23...
  • Page 128: Measurement Setup

    LTE Signal Analyzer Measurement Setup 1. Press the Measurements main menu key. 2. Press the LTE Summary submenu key to activate and display the LTE measurements in table format. Figure 5-16. LTE Summary 5-24 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 129: Lte Menus

    Spectral Back Spectral Emission List Emission Mask Category Page B (Opt. 1) Back Select Mask Page (Auto) Down Summary Bottom Back List Summary Table Back Figure 5-17. LTE Menu Layout (1 of 3) 3GPP MG PN: 10580-00234 Rev. Y 5-25...
  • Page 130: Lte Menus (Continued)

    Test Measurement Parameters RSRQ Setup EMF Units Back Point dBm/m V/m W/m Distance/Time SINR Setup Limits Back 20.0 dBm/m Back Back Back Start Data Collection Figure 5-18. LTE Menu Layout (2 of 3) 5-26 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 131: Lte Menus (Continued)

    LTE Signal Analyzer 5-8 LTE Menus LTE Menus (continued) Carrier Aggregation Center Freq 1.000 GHz Signal Standard Channel Set CF To Closest Channel 10 MHz Back Back Figure 5-19. LTE Menu Layout (3 of 3) 3GPP MG PN: 10580-00234 Rev. Y 5-27...
  • Page 132: Freq (Frequency) Menu

    Frequency History: Opens a list box that displays the last five selected frequencies. When a frequency is entered using the Center Frequency submenu key or the Signal Standard/Channel submenu keys, the list will be updated. Frequency History Figure 5-20. LTE Freq Menu 5-28 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 133: Standard List Menu

    Page Press Esc to close the Signal Standards list and return to the Frequency Down menu. Bottom List Figure 5-21. LTE Signal Standards Figure 5-22. LTE Signal Standard List, All and Favorites 3GPP MG PN: 10580-00234 Rev. Y 5-29...
  • Page 134: Amplitude Menu

    Auto Range is automatically turned Off. Units: Press this submenu key to set the units for all measurements and summary tables in either dBm or Watts. Units External Loss External Gain Backspace Figure 5-23. LTE Amplitude Menu 5-30 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 135: Setup Menu

    Ext. Ref. Freq (Only Some Models): Press to set the external reference frequency. The External Reference Frequency list opens. Highlight the desired frequency and press Enter. Back Figure 5-24. LTE Setup Menu 3GPP MG PN: 10580-00234 Rev. Y 5-31...
  • Page 136: Measurements Menu

    Save Measurement: Press this submenu key to save a measurement. You Summary may accept the default filename or enter your own filename. Refer to your Save instrument’s User Guide for information on file management functions. Measurement Figure 5-25. LTE Measurements Menu 5-32 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 137: Channel Power Occupied Bw Lower Adjacent Ch Pwr

    Spectral Emission Mask: The Spectral Emission Mask measurement supports the testing for “Operating Band Unwanted Emissions” described in the 3GPP TS 36.141 Base Station Conformance testing document. There is support for Category A and Category B (Option 1 only) masks which are...
  • Page 138: Modulation Menu

    PBCH Power Note: 256-QAM modulation scheme is supported only when Option 886 is installed. Back: Press this submenu key to return to the “Measurements Menu” on page 5-32. Figure 5-27. LTE Modulation Menu 5-34 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 139: Power Vs. Resource Block Menu

    RB Color Map Range: Press to adjust the RB Color Range manually. 0.0 dBm Back: Press this submenu key to return to the “Modulation Menu” on RB Color Map page 5-34. Range 100 dB Back Figure 5-28. LTE Power vs. Resource Block Menu 3GPP MG PN: 10580-00234 Rev. Y 5-35...
  • Page 140: Constellation Menu

    Values for the following measurements are displayed in a table beneath the graph: Ref. Signal (RS) Power EVM (rms) Freq Error Carrier Frequency Sync Signal (SS) Power EVM (pk) Freq Error (ppm) Cell ID Figure 5-29. LTE Constellation Menu 5-36 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 141: Control Channel Power Menu

    Ng (1/6): Determines the number of PHICH groups in a LTE sub-frame. Control Channel Back: Press this submenu key to return to the “Modulation Menu” on page 5-34. (1/6) Back Figure 5-30. LTE Control Channel Menu 3GPP MG PN: 10580-00234 Rev. Y 5-37...
  • Page 142: Over-The-Air Menu

    NA will be displayed until a valid Back measurement is achieved. Back: Press this submenu key to return to the “Measurements Menu” on page 5-32. Figure 5-31. LTE Over-the-Air Menu 5-38 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 143: Ota Scanner Menu

    Note: GPS coordinates are saved in addition to the measurement results, if GPS is on and a fix is established. Back: Press this submenu key to return to “Over-the-Air Menu” on page 5-38. Figure 5-32. LTE OTA Scanner Menu 3GPP MG PN: 10580-00234 Rev. Y 5-39...
  • Page 144: Over-The-Air Tx Test Menu

    Antenna Status indicators are not displayed. When set to On, the Antenna Status and PBCH Modulation Results table are displayed. Back: Press this submenu key to return to “Over-the-Air Menu” on page 5-38. Show Mod Results Back Figure 5-33. LTE Over-the-Air Tx Test Menu 5-40 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 145: Coverage Mapping Menu

    Distance/Time Setup settings. A running count of collected data points is displayed at the bottom of the screen. Press again to stop data collection. Measurement Setup Point Distance/Time Setup Back Start Data Collection Figure 5-34. Coverage Mapping Menu 3GPP MG PN: 10580-00234 Rev. Y 5-41...
  • Page 146: Carrier Aggregation Menu

    Component Carrier, if it is on. TAE: Time Alignment Error (TAE) is a measurement of the differential transmission delay between the Component Carriers. The 3GPP specification identifies the limits of the delays for the system to operate properly.
  • Page 147: Component Carrier Menu

    Back: Returns to the previous menu. To Closest Channel 10 MHz Back Figure 5-36. Component Carrier Menu EMF Menu Key Sequence: Measurements > Over-the-Air > EMF Refer to “LTE/TD-LTE EMF Menu” on page 8-2. 3GPP MG PN: 10580-00234 Rev. Y 5-43...
  • Page 148: Pass/Fail Test Menu

    Reset: Press this submenu key to restart the measurement or to begin a new pass/fail test measurement. Back Back: Press this submenu key to return to “Measurements Menu” on page 5-32. Figure 5-37. LTE Pass Fail Mode Menu 5-44 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 149: Marker Menu

    Marker Edit Markers Off: Turns off all markers. Reference Delta Back: Press this submenu key to return to “Measurements Menu” on page 5-32. Peak Search Markers Off Back Figure 5-38. LTE Marker Menu 3GPP MG PN: 10580-00234 Rev. Y 5-45...
  • Page 150: Power Vs Resource Block Marker Menu

    This key has Trigger no function when the instrument is in continuous sweep mode. Sweep Figure 5-40. LTE Sweep Menu 5-16 Measure Menu Displays the “Measurements Menu” on page 5-32. 5-46 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 151: Trace Menu

    This menu is not available in LTE measurement mode. 5-18 Limit Menu This menu is not available in LTE measurement mode. 5-19 Other Menus Preset, File, Mode and System are described in the instrument User Guide. 3GPP MG PN: 10580-00234 Rev. Y 5-47...
  • Page 152 5-19 Other Menus LTE Signal Analyzer 5-48 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 153: Introduction

    6. (Not applicable for all models) Press the Ext. Ref. Freq submenu key to configure the instrument using an external reference frequency. The External Reference Frequency list opens. Highlight the desired frequency and press Enter. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 154: Total Frame Pwr

    DwPTS Power The power of the Downlink Pilot Time Slot in Sub-Frame 1 in dBm or Watts. Transmit Off Power The mean power measured over a 70 μs window in the transmitter OFF period. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 155: Timing Error

    The Spectral Emission Mask (SEM) measurement supports the testing for “Operating Band Unwanted Emissions” described in the 3GPP TS 36.141 Base Station Conformance testing document. There is support for Category A and Category B (Option 1 only) masks which are automatically selected based on the current carrier frequency/channel and BW values.
  • Page 156: Rf Measurement Setups

    4. Press the Channel Spectrum submenu key again to select the Channel Spectrum menu, where you can adjust the span. Figure 6-2. Channel Spectrum 5. Press the Back submenu key to return to the RF menu. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 157 4. Press the Power vs. Time submenu key again to set the View and Sub-Frame Number parameters. Figure 6-3. Power vs. Time 5. Press the Back submenu key to return to the RF menu. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 158 2. Press the RF submenu key. 3. Press the ACLR submenu key to activate ACLR display and measurement (Figure 6-4). Figure 6-4. Adjacent Channel Leakage Ratio 4. Press the Back submenu key to return to the Measurements menu. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 159 5. To view the measurement data in a table format, press the Summary Table submenu key so On is underlined. Figure 6-5. Spectral Emission Mask 6. Press the Back submenu key to return to the RF menu. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 160 3. Press the RF Summary submenu key to activate and display the RF measurements in table format (Figure 6-6). Figure 6-6. RF Summary 4. Press the Back submenu key to return to the Measurements menu. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 161: Td-Lte Modulation Measurements

    LTE carrier uses 256-QAM modulation scheme, the corresponding EVM measurement result is displayed, provided Option 886 is installed and activated on your instrument (see Figure 6-7 on page 6-12). Cell ID Cell identifying information sent by the transmitter in the sync signal. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 162: Constellation

    Reset and the second number is the current measured value. Freq Error (ppm) Freq Error displayed in parts per million (ppm). Cell ID Cell identifying information sent by the transmitter in the sync signal. 6-10 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 163: Control Channel Power

    Modulation Summary displays the Ref Signal (RS) Power, Sync Signal (SS) Power, EVM (rms), Freq Error (Hz and ppm), Cell ID and PBCH Power in a table format. Refer to individual Modulation measurement descriptions for additional detail on each measurement. 3GPP MG PN: 10580-00234 Rev. Y 6-11...
  • Page 164: Modulation Measurement Setups

    This is useful in the monitoring of changes in traffic loading on the Note base station. To measure the power levels of the physical layer Control Channels, use the “Control Channel Power” measurement. 6-12 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 165 Constellation Measurement View Support for 256-QAM modulation as shown in the data legend above is available Note only with Option 886. 5. Press the Back submenu key to return to the Modulation menu. 3GPP MG PN: 10580-00234 Rev. Y 6-13...
  • Page 166 6-9). Press the key again to view or change the Display Mode and Control Channel EVM settings. Figure 6-9. Control Channel Power 4. Press the Back submenu key to return to the Modulation menu. 6-14 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 167 4x4 MIMO, the display may look similar to Figure 6-11. 4. Press the Back submenu key to return to the Measurements menu. Figure 6-10. Tx Time Alignment Figure 6-11. Tx Time Alignment 3GPP MG PN: 10580-00234 Rev. Y 6-15...
  • Page 168 3. Press the Modulation Summary submenu key to activate and display the modulation measurements in table format (Figure 6-12). Figure 6-12. Modulation Summary 4. Press the Back submenu key to return to the Measurements menu. 6-16 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 169: Td-Lte Over-The-Air (Ota) Measurements

    RSRP and RSSI measured in dB. SINR Signal-to-Inference Noise Ratio in dB. Dominance Dominance is the ratio of the power for the largest signal to the sum of all other signals found, in dB. 3GPP MG PN: 10580-00234 Rev. Y 6-17...
  • Page 170: Mapping (Coverage Mapping)

    TX antennas to two RX antennas on the same frequency and time, but different reference signals. Your Anritsu test instrument supports 2x2 and 4x4 MIMO configurations. Note that 4x4 MIMO may not be available if your instrument is loaded with older firmware. Refer to your instrument User Guide for instructions on updating firmware.
  • Page 171: Emf (Option 444 Only)

    This is the frequency error displayed in parts per million. Time Alignment Error (TAE) is a measurement of the differential transmission delay between the Component Carriers. The 3GPP specification identifies the limits of the delays for the system to operate properly.
  • Page 172 (Figure 6-14 on page 6-21). Press the Tx Test key again to display the OTA Tx Test menu. 4. Press the Show Mod Results key to display or hide the PBCH Modulation Results (Strongest SS) table. 6-20 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 173 Note Some instruments require the TD-LTE demod option to toggle this submenu to On. Figure 6-14. Over-the-Air Tx Test Measurement Mapping Refer to Chapter 7, “LTE and TD-LTE Coverage Mapping”. 3GPP MG PN: 10580-00234 Rev. Y 6-21...
  • Page 174 6-5 TD-LTE Over-the-Air (OTA) Measurements TD-LTE Signal Analyzer Carrier Aggregation Your Anritsu test instrument supports single-input single-output (SISO) as well as 2x2 and 4x4 multiple-input multiple-output (MIMO) systems. If a single antenna is active, the Component Carrier measurements illustrated in...
  • Page 175: Pass/Fail Tests

    Software Tools (MST) User Guide for creating new Pass/Fail tests. b. Press the Reset submenu key to erase a previous measurement/s and restart the Pass/Fail test. Figure 6-16. Past/Fail Test 3. Press the Back submenu key to return to the Measurements menu. 3GPP MG PN: 10580-00234 Rev. Y 6-23...
  • Page 176: Td-Lte Summary

    Physical Control Format Indicator Channel Power. This channel provides information to enable the UE to decode the PDSCH (Physical Downlink Shared Channel). Spectral Emission Mask Displays a Pass/Fail status for the signal measured against the selected mask. 6-24 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 177: Measurement Setup

    6-7 TD-LTE Summary Measurement Setup 1. Press the Measurements main menu key. 2. Press the TD-LTE Summary submenu key to activate and display the TD-LTE measurements in table format. Figure 6-17. TD-LTE Summary 3GPP MG PN: 10580-00234 Rev. Y 6-25...
  • Page 178: Td-Lte Menus

    Emission Mask Select Mask Trigger Source Page No Trig (Auto) Trigger Polarity Page Back Rising Falling Down Summary Summary Table Bottom Back List Back Back Figure 6-18. TD-LTE Menu Layout (1 of 3) 6-26 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 179: Td-Lte Menus (Continued)

    Back Measurement Parameters RSRQ Setup EMF Units Back Back Point dBm/m V/m W/m Distance/Time SINR Setup Limits 20.0 dBm/m Back Back Back Start Data Collection Figure 6-19. TD-LTE Menu Layout (2 of 3) 3GPP MG PN: 10580-00234 Rev. Y 6-27...
  • Page 180: Td-Lte Menus (Continued)

    6-8 TD-LTE Menus TD-LTE Signal Analyzer TD-LTE Menus (continued) Carrier Aggregation Center Freq 1.000 GHz Signal Standard Channel Set CF To Closest Channel 10 MHz Back Back Figure 6-20. TD-LTE Menu Layout (3 of 3) 6-28 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 181: Freq (Frequency) Menu

    Frequency History: Opens a list box that displays the last five selected frequencies. When a frequency is entered using the Center Frequency Frequency submenu key or the Signal Standard/Channel submenu keys, the list will be updated. History Figure 6-21. TD-LTE Freq Menu 3GPP MG PN: 10580-00234 Rev. Y 6-29...
  • Page 182: Standard List Menu

    Page Press Esc to close the Signal Standards list and return to the Frequency Down menu. Bottom List Figure 6-22. TD-LTE Signal Standards Figure 6-23. TD-LTE Signal Standard List, All and Favorites 6-30 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 183: Amplitude Menu

    Auto Range is automatically turned Off. Units: Press this submenu key to set the units for all measurements and summary tables in either dBm or Watts. Units External Loss External Gain Backspace Figure 6-24. TD-LTE Amplitude Menu 3GPP MG PN: 10580-00234 Rev. Y 6-31...
  • Page 184: Setup Menu

    Trigger: Press this key to open the “Trigger Menu” on page 6-33. Trigger Figure 6-25. TD-LTE Setup Menu 6-32 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 185: Sync Menu

    Trigger Polarity: Select the desired edge of the external trigger source to Trigger Polarity use for reference. Rising Falling Back: Press this submenu key to return to the “Setup Menu” on page 6-32. Back Figure 6-27. TD-LTE Sync Menu 3GPP MG PN: 10580-00234 Rev. Y 6-33...
  • Page 186: Measurements Menu

    Save Measurement: Press this submenu key to save a measurement. You Summary may accept the default filename or enter your own filename. Refer to your Save instrument’s User Guide for information on file management functions. Measurement Figure 6-28. TD-LTE Measurements Menu 6-34 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 187 6-37. The Spectral Emission Mask measurement supports the Emission Mask testing for “Operating Band Unwanted Emissions” described in the 3GPP TS 36.141 Base Station Conformance testing document. There is support for Category A and Category B (Option 1 only) masks which are automatically selected based on the current carrier frequency/channel and BW values.
  • Page 188: Channel Spectrum Menu

    Sub-Frame Number: If Sub-Frame is selected, press to enter the desired View sub-frame to view. Frame Sub-Frame Back: Returns to the “RF Menu” on page 6-35. Sub-Frame Number Back Figure 6-31. TD-LTE Power vs. Time Menu 6-36 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 189: Spectral Emission Menu

    Select Mask Summary Table: Displays the Spectral Emission measurement data in table form. (Auto) Back: Returns to the “RF Menu” on page 6-35. Summary Table Back Figure 6-32. TD-LTE Spectral Emission Menu 3GPP MG PN: 10580-00234 Rev. Y 6-37...
  • Page 190: Modulation Menu

    Cell ID PBCH Power Note: 256-QAM modulation scheme is supported only when Option 886 is installed. Back: Press this submenu key to return to “Measurements Menu” on page 6-34. Figure 6-33. TD-LTE Modulation Menu 6-38 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 191: Power Vs. Resource Block Menu

    RB Color Map Range: Press to adjust the RB Color Range manually. 0.0 dBm Back: Press this submenu key to return to the “Modulation Menu” on RB Color Map page 6-38. Range 100 dB Back Figure 6-34. TD-LTE Power vs. Resource Block Menu 3GPP MG PN: 10580-00234 Rev. Y 6-39...
  • Page 192: Constellation Menu

    Values for the following measurements are displayed in a table beneath the graph: Ref. Signal (RS) Power EVM (rms) Freq Error Carrier Frequency Sync Signal (SS) Power EVM (pk) Freq Error (ppm) Cell ID Figure 6-35. TD-LTE Constellation Menu 6-40 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 193: Control Channel Power Menu

    Ng (1/6): Determines the number of PHICH groups in a LTE sub-frame. Control Channel Back: Press this submenu key to return to the “Modulation Menu” on page 6-38. (1/6) Back Figure 6-36. TD-LTE Control Channel Menu 3GPP MG PN: 10580-00234 Rev. Y 6-41...
  • Page 194: Over-The-Air Menu

    EMF Measurement mode. Once the EMF Measurement mode is active, this button opens the “LTE/TD-LTE EMF Menu” on page 8-2. Back: Press this submenu key to return to the “Measurements Menu” on page 6-34. Back Figure 6-37. TD-LTE Over-the-Air Menu 6-42 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 195: Ota Scanner Menu

    Note: GPS coordinates are saved in addition to the measurement results, if GPS is on and a fix is established. Back: Press this submenu key to return to the “Over-the-Air Menu” on page 6-42. Figure 6-38. TD-LTE OTA Scanner Menu 3GPP MG PN: 10580-00234 Rev. Y 6-43...
  • Page 196: Over-The-Air Tx Test Menu

    Status and PBCH Modulation Results table are displayed. Back: Press this submenu key to return to the “Over-the-Air Menu” on page 6-42. Show Mod Results Back Figure 6-39. TD-LTE Over-the-Air Tx Test Menu 6-44 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 197: Coverage Mapping Menu

    Distance/Time Setup settings. A running count of collected data points is displayed at the bottom of the screen. Press again to stop data collection. Measurement Setup Point Distance/Time Setup Back Start Data Collection Figure 6-40. Coverage Mapping Menu 3GPP MG PN: 10580-00234 Rev. Y 6-45...
  • Page 198: Carrier Aggregation Menu

    Component Carrier, if it is on. TAE: Time Alignment Error (TAE) is a measurement of the differential transmission delay between the Component Carriers. The 3GPP specification identifies the limits of the delays for the system to operate properly.
  • Page 199: Component Carrier Menu

    Back: Returns to the previous menu. To Closest Channel 10 MHz Back Figure 6-42. Component Carrier Menu EMF Menu Key Sequence: Measurements > Over-the-Air > EMF Refer to “LTE/TD-LTE EMF Menu” on page 8-2. 3GPP MG PN: 10580-00234 Rev. Y 6-47...
  • Page 200: Pass/Fail Test Menu

    Reset: Press this submenu key to restart the measurement or to begin a new pass/fail test measurement. Back Back: Press this submenu key to return to the “Measurements Menu” on page 6-34. Figure 6-43. TD-LTE Pass Fail Mode Menu 6-48 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 201: Back: Press This Submenu Key To Return To The Page

    Marker Edit Markers Off: Turns off all markers. Reference Delta Back: Press this submenu key to return to the “Measurements Menu” on page 6-34. Peak Search Markers Off Back Figure 6-44. TD-LTE Marker Menu 3GPP MG PN: 10580-00234 Rev. Y 6-49...
  • Page 202: Power Vs Resource Block Marker Menu

    Trigger Sweep: Pressing this submenu key causes the instrument to make a single sweep when the instrument is in single sweep mode. This key has Trigger no function when the instrument is in continuous sweep mode. Sweep Figure 6-46. TD-LTE Sweep Menu 6-50 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 203: Measure Menu

    This menu is not available in TD-LTE measurement mode. 6-18 Limit Menu This menu is not available in TD-LTE measurement mode. 6-19 Other Menus Preset, File, Mode and System are described in the User Guide. 3GPP MG PN: 10580-00234 Rev. Y 6-51...
  • Page 204 6-19 Other Menus TD-LTE Signal Analyzer 6-52 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 205: Introduction

    Introduction Coverage Mapping allows users to measure and map Sync Signal Power, Reference Signal Received Power (RSRP), Reference Signal Receive Quality (RSRQ), and SINR. The Anritsu easyMap Tools program creates special maps compatible with Anritsu handheld spectrum analyzers. The software creates files with or without GPS information. The files compatible with coverage mapping will have a .map extension.
  • Page 206 (+) with the touchscreen. On instruments that do not have a touchscreen, use the arrow keys. Previously saved locations are displayed as colored squares. Current Location Set with Touchscreen or Arrow Keys Measurement Setup Figure 7-2. Indoor Coverage Mapping (GPS Off) PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 207: Anritsu Easymap Tools

    “Save the Coverage Mapping Information” on page 7-7. Anritsu easyMap Tools Anritsu easyMap Tools allows you to capture maps of any location and create Anritsu Map Files. These maps are viewed on the Anritsu instrument during coverage mapping. There are two Anritsu Map File formats: •...
  • Page 208: Instrument Settings

    7-3 Coverage Mapping LTE and TD-LTE Coverage Mapping Instrument Settings Setup 1. Create the appropriate map with Anritsu easyMap Tools. Refer to “Anritsu easyMap Tools” on page 7-3 and the software Help. Outdoor mapping requires a GeoEmbedded map or the default grid.
  • Page 209 Recall the Default Grid The instrument can make coverage mapping measurements even when an Anritsu easyMap Tools file of the current indoor or outdoor location is not available. In such cases, use the default grid map, save the KML points, and recall them at a later time with a map.
  • Page 210: Measurement Setup

    2. Press the Stop Data Collection main menu key to end the measurement process. Save the collected data as a .kml file, a tab-delimited text file (.mtd) or a .jpg file. Refer to “Save the Coverage Mapping Information” on page 7-7. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 211: Save The Coverage Mapping Information

    1. Connect your computer or mobile device to the instrument via the Web Remote Control server. To do this, enter the instrument IP address in the address bar of your HTML-5 compatible Web browser. Anritsu strongly recommends using Google Chrome, as other browsers do not fully support HTML-5.
  • Page 212 3. Click the map file name in the File column. Alternatively, you can select the checkbox next to the .kml file name, then click the Download button. 4. Click Open or Save in the pop-up dialog. Figure 7-4. Web Remote Control Window - File List Tab PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 213 Points. In the Save dialog, change the file name as appropriate, then press Enter. A tab delimited text file (.mtd) of the coverage mapping data currently displayed on the screen will be saved to the default location. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 214 Enter. A .jpg file of the current screen is saved to the default location. Figure 7-6. Coverage Mapping Measurement Saved as a .jpg File 7-10 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 215: Coverage Mapping Menus

    -117.0 dBm -117.0 dBm -9.0 dB 17.0 dB Poor Poor Poor Poor > > > > -120.0 dBm -120.0 dBm -12.0 dB 14.0 dB Back Back Back Back Back Figure 7-7. Coverage Mapping Menu 3GPP MG PN: 10580-00234 Rev. Y 7-11...
  • Page 216: Coverage Mapping Menu

    Point Distance/Time Setup. A running count of collected data points is Measurement displayed at the bottom of the map. Press this key again to stop data collection. Setup Point Distance/Time Setup Back Start Data Collection Figure 7-8. Coverage Mapping Menu 7-12 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 217: Mapping Save/Recall Menu

    Back Back: Returns to the “Coverage Mapping Menu” on page 7-12. Recall Sort By Name Type Date Sort Order Ascend Descend File Type Refresh Directories Figure 7-9. Mapping Save/Recall Menu 3GPP MG PN: 10580-00234 Rev. Y 7-13...
  • Page 218: Measurement Setup Menu

    Use the arrows, knob, or numeric keypad to change its value. Press Enter. Press Back to return to the Measurement Setup Menu. Back: Returns to the “Coverage Mapping Menu” on page 7-12. Figure 7-10. Measurement Setup Menu 7-14 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 219: Point Distance/Time Setup Menu

    Distance Units: Toggles the unit of measure between meters and feet. Delete Delete ALL Points: Deletes any and all map points. Points Back: Returns to the “Coverage Mapping Menu” on page 7-12. Back Figure 7-11. Point Distance/Time Setup Menu 3GPP MG PN: 10580-00234 Rev. Y 7-15...
  • Page 220 7-5 Coverage Mapping Menu LTE and TD-LTE Coverage Mapping 7-16 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 221: Introduction

    Option 444 adds the EMF Measurement menu to the Over-the-Air LTE and TD-LTE measurement modes. It must be used in conjunction with the Anritsu isotropic antenna, at a frequency range that is within specification of the instrument and antenna used. Refer to the isotropic antenna and signal analyzer technical data sheets.
  • Page 222: Lte/Td-Lte Emf Menu

    Key Sequence: Measurements > Over-the-Air > EMF Measurement On/Off: Starts the EMF Measurement and removes access to all other menu buttons. The measurement turns On only if the Center Frequency is set within the valid range and the Anritsu Isotropic Antenna is Measurement connected.
  • Page 223: Meas Params Menu (Lte/Td-Lte)

    Extr Factor On/Off: Toggles On/Off the Extrapolation Factor. The extrapolation factor, if turned On, displays the extrapolated field strength number for the selected capture bandwidth. Back: Returns to the “LTE/TD-LTE EMF Menu” on page 8-2. Back Figure 8-3. LTE/TD-LTE Meas Params Menu 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 224: Rs Display Menu

    RS Total Avg: Selects the display of Total Avg, which is the running average RS Total Avg for the entire measurement period. Back: Returns to the “Meas Params Menu (LTE/TD-LTE)” on page 8-3. Back Figure 8-4. RS Display Menu PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 225: P-Ss Display Menu

    P-SS Total Avg: Selects the display of Total Avg, which is the running P-SS Total Avg average for the entire measurement period. Back: Returns to the “Meas Params Menu (LTE/TD-LTE)” on page 8-3. Back Figure 8-5. P-SS Display Menu 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 226: S-Ss Display Menu

    S-SS Total Avg: Selects the display of Total Avg, which is the running S-SS Total Avg average for the entire measurement period. Back: Returns to the “Meas Params Menu (LTE/TD-LTE)” on page 8-3. Back Figure 8-6. S-SS Display Menu PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 227: Measurement Results (Lte/Td-Lte)

    Pass or Fail status at the end of the measurement period (Measurement Time × Number of Measurements). Field Strength (Avg) is the running average for the current Measurement Time. Field Strength (Total Avg) is the running average for all measurements completed thus far within the measurement period. 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 228 “--” indicates an invalid result. A maximum of six cell IDs can be detected. The Total row sums the isotropic numbers for the selected display parameter across Cell IDs. PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 229: Pass/Fail

    Test Status is marked as Pass in green. In the example in Figure 8-8, the extrapolated Field Strength (Ex Avg) is 22.22 mV/m. If all of the measurements pass, the Final Test Status is updated to Pass in green. Figure 8-8. LTE/TD-LTE EMF Measurement Display 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 230: W-Cdma Emf Menu

    Key Sequence: Measurements > OTA > EMF Measurement On/Off: Starts the EMF Measurement and removes access to all other menu buttons. The measurement turns On only if the Center Frequency is set within the valid range and the Anritsu Isotropic Antenna is Measurement connected.
  • Page 231 (default 6 V/m) for the test to pass. Back: Returns to the previous menu. Measurement Parameters EMF Units Limits 6.00 V/m Back Figure 8-10. W-CDMA EMF Menu (2 of 2) 3GPP MG PN: 10580-00234 Rev. Y 8-11...
  • Page 232: Meas Params Menu (W-Cdma)

    (system setup/power ratio) and/or per regulatory requirements (margins). Refer to “Measurement Results (W-CDMA)” on page 8-14. Back: Returns to the “W-CDMA EMF Menu” on page 8-10. Figure 8-11. W-CDMA Meas Params Menu 8-12 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 233: Display Menu

    Avg/Field Str Max/Field Str Avg/Field Str Min/Field Str Min/Field Str Total Avg/Field Str Back: Returns to the “Meas Params Menu (W-CDMA)” on page 8-12. Total Avg/Field Str Back Figure 8-12. W-CDMA Display Menu 3GPP MG PN: 10580-00234 Rev. Y 8-13...
  • Page 234: Measurement Results (W-Cdma)

    (rss) value of the common pilot signal recorded in each pcpich axis and k is the extrapolation factor provided by the network operator. For example: × √k max (pcpich total max) pcpich total max 8-14 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 235: Pass/Fail

    Pass in green. In the example in Figure 8-14, the Field Strength (Avg Meas) is 5.05 mV/m. If all of the measurements pass, the Final Test Status is updated to Pass in green. Figure 8-14. W-CDMA EMF Measurement Display 3GPP MG PN: 10580-00234 Rev. Y 8-15...
  • Page 236 8-4 W-CDMA EMF Menu EMF (Option 444) 8-16 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 237: A-1 Introduction

    Appendix A — Error Messages Introduction This Appendix provides a list of 3GPP error messages. Self Test and General Operation error messages are in the User Guide. 3GPP Messages Warning Messages 1. External Reference not found. Internal reference Locked successfully This message is displayed when the instrument has detected an external reference but couldn’t lock to the reference.
  • Page 238: A-3 Lte Messages

    2. No Trigger found When Trigger is set to ‘External’ and no external trigger is found, this message is displayed. Results that depend on the trigger are cleared (‘--’ is seen in the result area) PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 239 ... . 1-3 Anritsu contact ....1-4 freq error ..2-8, 3-3, 4-7, 5-9, 6-10 Anritsu easyMap Tools .
  • Page 240 WCDMA/HSPA+ ... . . 3-41 sub-frame power ....6-2 Index-2 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 241 ....3-14 threshold ......3-3 3GPP MG PN: 10580-00234 Rev. Y...
  • Page 242 Index-4 PN: 10580-00234 Rev. Y 3GPP MG...
  • Page 244: Morgan Hill, Ca

    Anritsu Company 490 Jarvis Drive Anritsu utilizes recycled paper and environmentally conscious inks and toner. Morgan Hill, CA 95037-2809 http://www.anritsu.com...

Table of Contents