Ma Measure Verification - Fluke 712 Calibration Manual

71x series process calibrators
Hide thumbs Also See for 712:
Table of Contents

Advertisement

71X Series
Calibration Manual
Input
Pressure
0 psi
300 psi
600 psi
900 psi
1200 psi
1500 psi
1200 psi
900 psi
600 psi
300 psi
*Vacuum is only specified for V1.3 and later. To display the firmware version, start with the 717 off, press
and hold M, then press the power button.
4. Carefully vent all pressure and disconnect the 717 from the deadweight tester.

mA Measure Verification

1. Press
2. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
3. Connect a test lead from Black AUX terminal of the 5520A to the 717 Calibrator
4. Connect a test lead from I- terminal of the Fluke 8508A to the 717 Calibrator mA
46
717 1500G
Display
Reading
0.8 to -0.8
300.8 to 299.3
600.8 to 599.3
900.8 to 899.3
1200.8 to 1199.3
1500.8 to 1499.3
1200.8 to 1199.3
900.8 to 899.3
600.8 to 599.3
300.8 to 299.3
Input
Pressure
0 psi
1000 psi
2000 psi
3000 psi
4000 psi
5000 psi
4000 psi
3000 psi
2000 psi
1000 psi
O
to turn power on. The display should read:
----------
mA
then change to:
0.000 mA
0.00PSI
terminal.
Com terminal.
terminal.
Input
Pressure
0 psi
600.0 psi
1200.0 psi
1800.0 psi
2400.0 psi
3000.0 psi
2400.0 psi
1800.0 psi
1200.0 psi
600.0 psi
717 5000G
2002.5 to 1997.5
3002.5 to 2997.5
4002.5 to 3997.5
5002.5 to 4997.5
4002.5 to 3997.5
3002.5 to 2997.5
2002.5 to 1997.5
717 3000G
Display
Reading
1.5 to -1.5
601.5 to 598.5
1201.5 to 1198.5
1801.5 to 1798.5
2401.5 to 2398.5
3001.5 to 2998.5
2401.5 to 2398.5
1801.5 to 1798.5
1201.5 to 1198.5
601.5 to 598.5
Display
Reading
2.5 to -2.5
1002.5 to 997.5
1002.5 to 997.5