718 And 718Ex Specifications - Fluke 712 Calibration Manual

71x series process calibrators
Hide thumbs Also See for 712:
Table of Contents

Advertisement

718 and 718Ex Specifications

Accuracy is specified for 1 year after calibration at operating temperatures of -10 °C to +
55 °C.
To display firmware version, start with the unit off, press and hold M, then press O.
Model
Range SI
718-1G
(-1 to 1) PSI
718-30G
(-12 to 30) PSI
718-100G
(-12 to 100) PSI
718-300G
(-12 to 300) PSI
Accuracy: Pressure Accuracy is +/- 0.05% of range
Temperature coefficient: .01 % of range per °C for temperature ranges of between −10 °C to 18 °C
and 28 °C to 55 °C.
Range
Range
718: Version 1.2
24 mA
718: Version 1.3
24 mA
718 EX
24 mA
Table 29. Pressure Specifications
For version 1.3 or higher
Range Metric
(-7 to 7) kPa
(-83 to 207) kPa
(-83 to 690) kPa
(-83 to 2068) kPa or 2.1 mPa
Table 30. Pressure Display, Pressure Module Input
Resolution
Refer to the Instruction Sheet for the pressure module
Table 31. DC mA Input
Resolution
0.001 mA
0.001 mA
0.001 mA
Max SI
5 PSI
60 PSI
200 PSI
375 PSI
Accuracy, ± ( % of Reading + Counts)
0.025 + 1
0.015 + 2
0.02 + 2
Process Calibrators
Specifications
Max Metric
34.5 kPa
413 kPa
1379 kPa or 1.4 mPa
2586 kPa or 2.6 mPa
Accuracy
19