5.6.1 Monitor Units
The Monitor Units parameter determines the units of the Control Input sensor to use
for creating the proportional voltage output. The Monitor Out scaling parameter set-
tings will be entered using the units chosen for this parameter.
Menu Navigation:
Output SetupQOutput 2QOutput Type (Voltage)QOutput Mode (Monitor Out)Q
Control Input (None, Input A, Input B)QMonitor Out Units (Kelvin, Celsius, or Sensor)
Default: Kelvin
Interface Command: ANALOG
5.6.1.1 Polarity and Monitor Out Scaling Parameters
In the Monitor Out and Open Loop modes, the voltage output can be configured as
either unipolar (0 V to +10 V) or bipolar (–10 V to +10 V). In bipolar mode, the Monitor
Out –10 V setting determines the temperature or sensor value at which the output
should be –10 V. In unipolar mode, the Monitor Out 0 V setting determines the tem-
perature or sensor value at which the output should be 0 V. The Monitor Out +10 V
setting determines the temperature or sensor value at which the output should be
+10 V in either unipolar or bipolar modes.
Bipolar
Input mode
Unipolar
For example, if Polarity is set to Bipolar, then setting the Monitor Out –10 V parameter
to 0 K and the Monitor Out +10 V parameter to 100 K will cause the voltage output to
correspond to the input temperature as shown in FIGURE 5-4. In this case if the actual
reading was 50 K, then the output would be at 0 V (middle of the scale).
Bipolar
FIGURE 5-4 Analog output with polarity set to bipolar
FIGURE 5-5 shows the output if the Polarity parameter is set to Unipolar. In this case if
the actual reading was 50 K, the voltage output would be +5 V (middle of the scale).
Unipolar
FIGURE 5-5 Output with polarity parameter set to unipolar
Menu Navigation:
Output SetupQOutput 2 QOutput Type (Voltage)QOutput Mode (Monitor Out)Q
Control Input (None, Input A, Input B)QMonitor Out Units (Kelvin, Celsius, or Sensor)
QPolarity (Unipolar or Bipolar)
ENTER
Output SetupQOutput 2 QMonitor Out
below)
Output SetupQ Output 2 QMonitor Out
Input
Lowest
Middle
–10 V
Output
Lowest
Middle
0 V
FIGURE 5-3 Unipolar and bipolar mode
Input
0 K
–10 V
Output
0 K
0 V
0 V
(Unipolar) or
+10 V
Q(See note below)
5.6.1 Monitor Units
Highest
0 V
+10 V
Input
Highest
+5 V
+10 V
Output
50 K
100 K
0 V
+10 V
Input
50 K
100 K
+5 V
+10 V
Output
-10 V
(Bipolar)Q(See note
|
www.lakeshore.com