Emul; Model 331/332 Emulation Mode Query; Filter; Filter? Input Filter Parameter Query - Lakeshore 335 User Manual

Table of Contents

Advertisement

EMUL

Model 331/332 Interface Emulation Mode Command
Input
EMUL <emulation mode>,<PID scaling mode>[term]
Format
nn
<emulation mode>
<PID scaling mode>
Remarks
The 331 and 332 emulation modes provide a means of using the Model 335 in place
of a Model 331 or 332 in a software controlled system without updating the software.
The emulation mode setting only affects remote operation; front panel operation of
the Model 335 is not changed. Please reference the Model 331 or 332 user's manual
for information on the interface commands for those models. For more information
on using the emulation modes, see section 5.11.
EMUL?
Model 331/332 Interface Emulation Mode Query
Input
EMUL? [term]
Returned
<emulation mode>,<PID scaling mode>[term]
Format
n,n (refer to command for description)

FILTER

Input Filter Parameter Command
Input
FILTER <input>,<off/on>,<points>,<window>[term]
Format
a,n,nn,nn
<input>
<off/on>
<points>
<window>
Example
FILTER B,1,10,2[term] filter input B data through ten readings with 2% of
full scale window.
FILTER?
Input Filter Parameter Query
Input
FILTER? <input>[term]
Format
a
<input>
Returned
<off/on >,<points>,<window>[term]
Format
n,nn,nn (refer to command for description)
HTR?
Heater Output Query
Input
HTR? <output>[term]
Format
n
<output>
Returned
<heater value>[term]
Format
±nnn.n
<heater value>
0=None (335), 1=331, 2=332
PID control scaling: 0=335 (Temperature), 1=331/332
(Sensor)
Specifies input to configure: A or B.
Specifies whether the filter function is 0 = Off or 1 = On.
Specifies how many data points the filtering function uses.
Valid range = 2–64.
Specifies what percent of full scale reading limits the filtering function.
Reading changes greater than this percentage reset the filter. Valid
range = 1 to 10%.
Specifies input to query: A or B.
Heater output to query: 1 = Output 1, 2 = Output 2
Percent of full scale current for Output 1 and Output 2 in
Current mode, or percent of full scale voltage for Output 2
in Voltage mode.
6.4.1 Interface Commands
|
www.lakeshore.com
115

Hide quick links:

Advertisement

Table of Contents
loading