Model 335 Temperature Controller
Remote Interface Mode Command
MODE <mode>[term]
n
<mode>
MODE 2[term] places the Model 335 into remote mode with local lockout.
Remote Interface Mode Query
MODE?[term]
<mode>[term]
n (refer to command for description)
Manual Output Command
MOUT <output>,<value>[term]
n, +nnnnn[term]
<output>
<value>
MOUT 1,22.45[term] Output 1 manual output is 22.45%.
Manual output only applies to outputs in Closed Loop PID, Zone, or Open Loop modes.
Manual Output Query
MOUT? <output>[term]
n
<output>
<value>
+ nnnnn[term] (refer to command for description)
Operational Status Query
OPST? [term]
<bit weighting> [term]
nnn
The integer returned represents the sum of the bit weighting of the operational sta-
tus bits. Refer to section 6.2.5.2 for a list of operational status bits.
Operational Status Enable Command
OPSTE <bit weighting> [term]
nnn
Each bit has a bit weighting and represents the enable/disable mask of the corre-
sponding operational status bit in the Operational Status Register. This determines
which status bits can set the corresponding summary bit in the Status Byte Register.
To enable a status bit, send the command OPSTE with the sum of the bit weighting for
each desired bit. Refer to section 6.2.5.2 for a list of operational status bits.
Operational Status Enable Query
OPSTE?[term]
<bit weighting> [term]
nnn (Refer to section 6.2.5.2 for a list of operational status bits)
0 = local, 1 = remote, 2 = remote with local lockout.
Specifies output to configure: 1 or 2.
Specifies value for manual output.
Specifies which output to query: 1 or 2.