Carescape V100 Vital Signs Monitor - GE CareScape V100 Service Manual

Vital sings monitor
Hide thumbs Also See for CareScape V100:
Table of Contents

Advertisement

D-YS & D-YSE
D-YS & D-YSPD
DS-100A
OXI-A/N (adult)
OXI-A/N (neonate)
OXI-P/I
Neonatal sensor accuracy
Sensor light source
Wavelength***
Power dissipation
*
The MAX-N, D-YS, OXI-A/N, and OxiCliq N were tested on patients >40 kg.
**
The accuracy specification has been determined between saturations of 80%-100%.
*** Information about wavelength range can be especially useful to clinicians.
2048723-002C
Technical specifications and default settings: Specifications
Specifications: Nellcor sensor accuracy

CARESCAPE V100 Vital Signs Monitor

± 3.5 digits
± 3.5 digits
± 3 digits
± 3 digits
± 4 digits
± 3 digits
When sensors are used on neonatal subjects as
recommended, the specified accuracy range is increased
by ±1 digit, as compared to adult usage, to account for the
theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood. For example, MAX-N
accuracy on neonates is ±3 digits, rather than ±2 digits.
Infrared: 890 nm (nominal)
Red: 660 nm (nominal)
Infrared: 22.5 mW (max)
Red: 30 mW (max)
A-9