EMPV Cleaning

1260 Nanoflow Pump EMPV Cleaning Description

Depending on the application, particles can sometimes collect in the EMPV.
This fast cleaning routine is designed to remove such particle deposits. The
routine should always be performed when the EMPV is suspected of being
leaky or contaminated with particles.
The outlet of the EMPV is plugged with an SST blank nut. After a short
flushing routine the EMPV is closed and the pressure is increased to
approximately 380 bar. The EMPV is then opened and the pressure is released
very quickly. This procedure is repeated several times in a sequence.

Running the Test

1 Fill vacuum degasser channel A1 and B1 with solvents (the test requires
2 Plug the pump outlet with blank nut at EMPV outlet.
3 Disconnect the EMPV to flow sensor at the EMPV outlet. Plug the EMPV
4 Execute test.
5 Check result with Pressure Test if necessary.
6 Remove the blank nut.
7 Reconnect the EMPV to flow sensor capillary. Do not overtighten!
1260 Nanoflow Pump User Manual
filled vacuum chambers). We recommend that you use channel A with
aqueous solvent. If you use a different channel, you must ensure
the miscibility of the solvent
that no precipitation of buffer occurs
outlet port with blank nut (01080-83202).
Test Functions and Calibration

EMPV Cleaning

8