Running The Leak Test - Agilent Technologies 1260 Infinity User Manual

Nanoflow pump
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

8
Test Functions and Calibration
Leak Test

Running the Leak Test

When
If problems with the pump are suspected
Tools required
Wrench 1/4 inch
Parts required
#
1
1
1
Preparations
• Place two bottles of LC-grade isopropyl alcohol in channels A2 and B2.
• The EMPV is not designed for pressure tightness towards the waste port. Install the manual
• Move the inline filter outlet capillary from the EMPV to to manual purge valve.
Make absolutely sure that all parts of the flow path that are part of the test are very
N O T E
thoroughly flushed with IPA before starting to pressurize the system! Any trace of other
solvents or the smallest air bubble inside the flow path definitely will cause the test to fail!
Running the test from the Agilent Lab Advisor software
1 Select the Leak Test from the Test Selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by slowly opening the purge valve when the test has
N O T E
finished.
"Evaluating the Results"
H I N T
leak test results.
For detailed instructions refer to the Agilent Lab Advisor software Tool.
H I N T
118
p/n
Description
G1313-87305
Restriction Capillary
01080-83202
Blank nut
500 ml Isopropanol
purge valve from the accessories kit to pump head A.
on page 119 describes the evaluation and interpretation of the
1260 Nanoflow Pump User Manual