Instructions for Use

Maintenance

Personnel responsible for pump operation and maintenance must be well-trained
and aware of the accident prevention rules.
• Death may result from contact with high voltages. Always take extreme care
WARNING
and observe the accident prevention regulations in force.
• When machine is powered up, be careful of moving parts and high voltages.
• If you have to perform maintenance on the pump after a considerable time in
operation, allow it to cool as the temperature of the outer surface may be in
excess of 60 °C.
• Always disconnect your power supply to the pump before beginning
maintenance work.

Cleaning

Exterior

The exterior surfaces of the IDP-10 may be cleaned with a cloth slightly moistened
with water. If is necessary alcohol or a mild detergent may be used instead. Do not
pour or spray any liquids onto the pump.

Interior

Run the pump periodically at atmosphere for a minute or two to flush it out.
NOTE
Cleaning the Pump
Unlike conventional oil-sealed pumps, Agilent dry scroll pumps do not contain fluid
for the cleansing of accumulated dust and debris. Run the pump periodically at
atmosphere for a minute or two to flush it out. Until experience is gained on your
specific process, flush the pump regularly and adjust this schedule according to
your specific conditions.
178
Agilent IDP-10 Dry Scroll Vacuum Pump User Manual