Evaluating the Results

The sum of all leaks between the pump and the blank nut will be indicated by a
pressure drop of >2 bar/minute at the plateau. Note that small leaks may cause
the test to fail, but solvent may not be seen leaking from a module.
Please notice the difference between an error in the test and a failure of the test! An error
N O T E
means that during the operation of the test there was an abnormal termination. If a test
failed, this means that the results of the test where not within the specified limits.
Often it is only a damaged blank nut itself (poorly shaped from overtightening) that causes
N O T E
a failure of the test. Before investigating on any other possible sources of failure make sure
that the blank nut you are using is in good condition and properly tightened!
1260 Nanoflow Pump User Manual
Test Functions and Calibration
Normal Mode Pressure Test
8
115