Device information

Manufacturer:
Sartorius
Model:
SECURA2102
Serial number:
0012345678
Version BAC:
00-50-02.xx CN:1701
Version APC:
01-70-02.xx CN:6587

Calibration/Adjustment

isoCAL
Calibration report
Calibration report
2014-10-04: 2 Protocols
2014-10-01: 2 Protocols
2014-09-25: 3 Protocols
2014-09-24: 3 Protocols
2014 09 20: 4 Protocols

Weighing

SQmin
Safety level
Ambient conditions
Application
Stability signal
6.2.3
Device Information
The manufacturer, model, serial number, and software version of your balance are
displayed here .
6.2.4
Calibration/Adjustment
With Secura
®
models, it is possible to set how the automatic calibration/adjustment
function isoCAL is started .
With Secura
and Quintix
models, saved reports about the most recent calibration
®
®
procedures, including deviations identified, can be viewed and output to a PC or
Sartorius lab printer .
isoCAL
menu option (Secura
Select the desired start option for isoCAL .
t
Off
: isoCAL is switched off .
Info, manual start
: A note appears on the display when predefined time
intervals or temperature values are exceeded . isoCAL can be started
manually .
Automatic
: isoCAL is switched on . Calibration/adjustment starts
automatically when predefined time intervals or temperature values are
exceeded .
Calibration report
menu option (Secura
Select the desired report by date .
t
If there is more than one calibration report per date, use the
navigate through the reports .
Select
a
to print the displayed report .
t
A maximum of 99 entries are saved per day .
Calibration reports are deleted after 30 days .
6.2.5
Weighing
A number of basic settings for the weighing functions can be defined here . Options
are limited for verified balances . All settings given in the menu are permitted .
SQmin
menu option (Secura
If the SQmin function is set up on the balance, the function can be switched
Off
here (see "SQmin Minimum Weighing (Secura
System Settings (Setup Menu)
®
only):
and Quintix
®
®
only):
®
only)" on page 44) .
User Manual Secura
only):
®
<
>
and
buttons to
On
or
31
®
, Quintix
®
, Practum
®

Hide quick links: