Error Message
Supply Pressure Too Low
Value out of range
Function not available
Illegal Operation
72
Table 8. Error Messages (cont.)
Explanation
Failed to create supply pressure for the
current setpoint.
Value input is out of range.
Selected function is not available for the
current state.
Illegal operation for the current state.
Potential Root Cause
1. The pressure volume is too large,
reduce the volume and try again.
2. The leakage of pressure volume is too
large. Test the leakage and try again.
3. Ambient pressure is lower than the
negative gauge setpoint. Change the
negative gauge setpoint to a reasonable
value at current ambient pressure.
4. The unit needs service.
Try to configure a valid value.