729/729FC
Users Manual
Caution
To prevent possible damage to the Product or
to equipment under test:
Never apply more than 10 lb-ft. of torque
between the pressure module fittings, or
between the fittings and the body of the
module.
Always apply correct torque between the
pressure module fitting and connecting
fittings or adapters.
Never apply pressure above the rated
maximum printed on the pressure module.
Use the pressure module only with
specified materials. See the printing on the
pressure module or the pressure module
instruction sheet for the acceptable
material compatibility.
52
Connect a pressure module to the Product as shown in
Figure 19. The pressure module shows on the display after
several seconds once installed.
The Product automatically senses the pressure module
attached and sets its range accordingly.
Before doing a task that sources or measures
pressure, zero the Product.
1.
With the pressure module highlighted in the third
screen, push  (Zero External) to zero the
pressure. When zeroing absolute pressure modules,
enter the current barometric pressure when the
displayed value turns to yellow highlight
2.
Zero the pressure. See the module's Instruction Sheet.
Modules can have different zeroing procedures that
depend on module type.