Performance Verification
5-8
With the power up settings in memory, you can initialize the 371B, as instructed,
by performing the following:
1. Press the setup RECALL button (with the memory location set to
location 1).
2. Check that the SET 1 message is displayed.
DISPLAY
Mode
VERTICAL
HORIZONTAL
INVERT
STEP GENERATOR
SOURCE
NUMBER OF STEPS
OFFSET
STEP MULTI .1X
STEP AMPLITUDE
INVERT
COLLECTOR SUPPLY
VARIABLE
PEAK POWER WATTS
POLARITY
MEASUREMENT
CURSOR
MEMORY
INDEX NUMBER
STORE
1 A/DIV
1 V/DIV
Off
CURRENT
2
0.00 mA
Off
50 nA
Off
0.0%
300 W
NPN+
REPEAT
OFF
1
371B Service Manual