Error Information
MCT Error Messages

WRONG VALVE

Error ID: 32130
Valve not supported in this module.
Probable cause
1
For G7116B:
N O T E
• G4232D (5067-4240: 2ps/10pt Valve 1300 bar)
• G4231C (5067-4241: 2ps/6pt Valve 1300 bar)
• G4234C (5067-4273: 6 Column Selector 1300 bar)
• G4239C (5067-4233: 8 Column Selector 1300 bar)
For G7116A:
• G4232C (5067-4283: 2ps/10pt Valve 800 bar)
• G4231A (5067-4282: 2ps/6pt Valve 800 bar)
• G4237A (5067-4279: 4 Column Selector 800 bar)
For G7116A Bio-inert:
• G5632A (5067-4132: 2ps/10pt Valve 600 bar, bio)
• G5631A (5067-4148: 2ps/6pt Valve 600 bar, bio)
• G5639A (5067-4134: 4 Column Selector 600 bar, bio)
98
A valve head has been installed which is
not usable with G7116A.
Suggested actions
Only valve heads up to 800 bar or with up to
four positions can be used in G7116A.
Exchange the valve head with a suitable one.
For changing the valve head follow the
instructions Replace Valve Heads.
1260 Infinity II Multicolumn Thermostat User Manual