7
Error Information
MCT Error Messages
MCT Error Messages
These errors are module specific.
VALVE INIT FAILED
Error ID: 32030
During initialization, the valve actuator turns until the encoder reads the
reference index position. An error is generated, if the reference index cannot
be found within a given time.
Probable cause
1
2
3
96
Defect in cable connection of valve drive
control.
Defect in cable connection of valve actuator
encoder reader.
Defect of valve drive or valve actuator
encoder reader.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
1260 Infinity II Multicolumn Thermostat User Manual