System description
4. System description
12
Measurement units of the portable glossmeter
family can be used to determine the gloss level of
paint coatings, plastics, ceramics and metal
surfaces.
Light is directed at the surface of the sample at a
defined angle and the reflected light is measured
photoelectrically (reflectometer).
Depending on the typical gloss level of the test
object, reflectometers that direct light onto the
surface at different angles (geometry) can be used.
Measurement units are equipped with standard
geometries of 20°, 60° or 85°. All three of these
geometries are integrated into the
Functions described in this manual in terms of
geometry selection are only available with the three
angle device.
In addition to measuring individual gloss values, it is
also possible to record, save and statistically
evaluate series of measurements consisting of up to
999 values.
The operate button and scroll wheel are used to
control the system. System operation is supported
by display messages (autodiagnosis and error
messages).
The measurement unit conforms to the standards
ISO 2813, ASTM D 523, ASTM D 2457, DIN 67530
and ISO 7668, JIS Z 8741 (excluding 60°S)
three angle