10
in Figure 4-5
TABLE 4-2
Cp-Ls ACCURACY BETWEEN
0.4 AND 4 ON THE CGRL DIAL.
Points to remember in measurements above 20
kHz are:
1. The Cp
-
L, bridges are more accurate than
the C,
-
L bridges.
2. Rccuracy is greater with the CGRL dial a t a
low setting, s a y between 0.4 and 4.
3.
While the basic 1% bridge accuracy may be
plus or minus, the error introduced above 20 kHz is
always minus.
For greater accuracy between 50 and
100 kHz, add 1% t o the indicated value.
4. When measuring D or Q above 20 kHz, always
use the C p
-
L, bridges.
The above information is given merely a s a guide
for those wondering what accuracy they might reason-
ably expect at frequencies from 20 to 100 kHz.
Frequency
50
kHz
100
kHz
NOTE
Bridges are not tested a t these fre-
quencies, and thus operation above 20
kHz is not included in the specifications.
Basic Bridge
Accuracy
*I
%
*I%
Limits of Error
Added Above 20 kHz
+0, -1%
+0, -2.5%
ACCURACY
4-6
Net Accuracy
Limits*
+I
%,
-2%
+I%, -3.5%