H82.0.01.6C-02 Operating Manual GMH 5130 page 3 of 14 _____________________________________________________ _____________________________________________________________________________ 1 General Note Read this document carefully and get used to the operation of the device before you use it. Keep this document within easy reach near the device for consulting in case of doubt.
_DANGER Product Specification 3.1 Scope of supply The scope of supply includes: GMH 5130 with 2 AAA batteries Operation manual Short form manual 3.2 Operation and maintenance advice 1. Battery operation: If ‘bAt’ is shown in the lower display the battery has been used up and needs to be replaced. However, the device will operate correctly for a certain time.
H82.0.01.6C-02 Operating Manual GMH 5130 page 5 of 14 _____________________________________________________ _____________________________________________________________________________ 4 Handling 4.1 Display elements Main display: measuring value Secondary display: display of minimum / maximum / memorized measuring value Arrows for selected measuring unit Rating of battery status...
H82.0.01.6C-02 Operating Manual GMH 5130 page 6 of 14 _____________________________________________________ _____________________________________________________________________________ 4.4 Pop-up clip Handling: • Pull at label “open” in order to swing open the pop-up clip. • Pull at label “open” again to swing open the pop-up clip further.
H82.0.01.6C-02 Operating Manual GMH 5130 page 7 of 14 _____________________________________________________ _____________________________________________________________________________ 5 Start Operation Connect sensors, turn device on via key. After segment test the device displays some configuration: if standard measurement is selected if fast measurement is selected if peak value detection is selected After that the device is ready for measuring.
H82.0.01.6C-02 Operating Manual GMH 5130 page 8 of 14 _____________________________________________________ _____________________________________________________________________________ Averaging Filter 1 … 120 Averaging period in seconds Averaging function is deactivated Auto Power-Off: Select power-off delay 1 ... 120 Power-off delay in minutes. Device will be automatically switched off as soon as this time has elapsed if no key is pressed/no interface communication takes place.
H82.0.01.6C-02 Operating Manual GMH 5130 page 9 of 14 _____________________________________________________ _____________________________________________________________________________ 7.2.1 Standard measuring (slow) Measuring rate 4 Hz, averaging and filter functions are active. Application: Measuring of slowly changing or static pressures, e.g. measuring of leakproofness, atmospheric pressure... Highest accuracy, high noise immunity, low power consumption.
H82.0.01.6C-02 Operating Manual GMH 5130 page 10 of 14 _____________________________________________________ _____________________________________________________________________________ 8 Output The output can be used as serial interface (for USB 5100 interface converter). If the output is not needed, it is strongly recommended to deactivate it (Out oFF) to lower power consumption.
H82.0.01.6C-02 Operating Manual GMH 5130 page 11 of 14 _____________________________________________________ _____________________________________________________________________________ 10 Pressure Connection to the Sensors The device is designed to be connected to the sensors of the GMSD/GMXD...-series without a new calibration being necessary. Therefore a great variety of replaceable sensors of e.g. -1.999 ... 2.500 mbar relative up to 0 ...
H82.0.01.6C-02 Operating Manual GMH 5130 page 12 of 14 _____________________________________________________ _____________________________________________________________________________ 13 Error and System Messages Display Description What to do? No sensor connected Switch off device and connect sensor Connected sensor or device defective If 2nd sensor available, check if device is ok. Return defective device/sensor to manufacturer for repair Err.9...
Need help?
Do you have a question about the GMH 5130 and is the answer not in the manual?
Questions and answers