Display Messages - McKesson TRUE METRIX PRO Instruction Manual

Professional monitoring blood glucose meter
Table of Contents

Advertisement

Messages
Display
Reason
Invalid
Hematocrit
Temperature
Error
Too Cold/
Too Hot
Sample Not
Detected or
Using Wrong
Test Strip
Used Test Strip,
Test Strip out-
side of vial too
long, Sample on
top of Test Strip.
Meter
Error
Action
Repeat with new test strip, using
capillary whole blood from the
finger or forearm or venous whole
blood collected with sodium
heparin, lithium heparin or EDTA.
If error persists, call for assistance.
Move meter and test strips
to area between 41°F-104°F;
wait 10 minutes for System
to reach room temperature
before testing.
Retest with new
McKesson TRUE METRIX®
PRO Professional
Monitoring Blood Glucose
Test Strip and larger sample.
Repeat with new test
strip. Make sure sample
is touched to edge of test
strip (not top). If error
persists, call for assistance.
assistance.
41
Call for