Do you have a question about the TRUE METRIX PRO and is the answer not in the manual?
Questions and answers
Dale Simms
January 17, 2025
I **** interested in your recommendation of cleaning products and disinfecting types. Currently we are using a wipe with 70% alcohol the a cavi wipe afterwards. We have had trouble with our glucometer shutting down, possibly from too much moisture. Also we start an IV and I would like to take the dot of blood from the IV angio- would this be acceptable with this glucometer? Thank you
1 comments:
Mr. Anderson
May 14, 2025
The recommended cleaning and disinfecting method for the McKesson PRO glucometer involves using wipes that have been tested for up to 10,950 cleaning and disinfecting cycles, equal to 10 cleanings per day over 3 years. This ensures the meter's proper function without moisture issues when done according to guidelines. Healthcare professionals must wear gloves while cleaning and disinfecting.
It is not acceptable to take a blood sample from an IV angio with this glucometer. The meter is validated for fresh capillary whole blood from the finger and forearm, and venous blood drawn in sodium heparin tubes—not from IV lines.
This answer is automatically generated
Anne Clark
March 7, 2025
how long after opening vial must control solution be discarded
1 comments:
Mr. Anderson
March 7, 2025
The control solution for the McKesson TRUE METRIX® PRO can be used for up to 3 months after first opening or until the expiration date printed on the label has passed, whichever comes first.
This answer is automatically generated
Sheila Sturm
March 19, 2025
what is a e-3 code
1 comments:
Mr. Anderson
May 14, 2025
An E-3 code for McKesson TRUE METRIX PRO indicates that the meter has detected the test strips are compromised.
Need help?
Do you have a question about the TRUE METRIX PRO and is the answer not in the manual?
Questions and answers
I **** interested in your recommendation of cleaning products and disinfecting types. Currently we are using a wipe with 70% alcohol the a cavi wipe afterwards. We have had trouble with our glucometer shutting down, possibly from too much moisture. Also we start an IV and I would like to take the dot of blood from the IV angio- would this be acceptable with this glucometer? Thank you
The recommended cleaning and disinfecting method for the McKesson PRO glucometer involves using wipes that have been tested for up to 10,950 cleaning and disinfecting cycles, equal to 10 cleanings per day over 3 years. This ensures the meter's proper function without moisture issues when done according to guidelines. Healthcare professionals must wear gloves while cleaning and disinfecting.
It is not acceptable to take a blood sample from an IV angio with this glucometer. The meter is validated for fresh capillary whole blood from the finger and forearm, and venous blood drawn in sodium heparin tubes—not from IV lines.
This answer is automatically generated
how long after opening vial must control solution be discarded
The control solution for the McKesson TRUE METRIX® PRO can be used for up to 3 months after first opening or until the expiration date printed on the label has passed, whichever comes first.
This answer is automatically generated
what is a e-3 code
An E-3 code for McKesson TRUE METRIX PRO indicates that the meter has detected the test strips are compromised.
This answer is automatically generated