Test Strips - McKesson TRUE METRIX PRO Instruction Manual

Professional monitoring blood glucose meter
Table of Contents

Advertisement

Test Strip
① Contact End - Insert into Test Port with contact
blocks facing up.
② Sample Tip - Touch Tip to top of drop of sample
(blood or control solution) after Drop Symbol
appears in the Display.
Note: Insert test strip into meter before touching Sample
Tip to blood or control solution drop.
Sample Placement
Correct
• Allow sample drop to be drawn into Sample Tip
until testing begins (meter beeps and dashes move
across Display) .
• Do not smear or scrape drop with test strip.
• Do not apply more sample to Sample Tip after testing
begins.
• Do not apply blood or control to top of test strip.
Do not insert Sample Tip with sample into Test Port.
May damage meter.
Top of Test Strip
2
10
1
Incorrect

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TRUE METRIX PRO and is the answer not in the manual?

Questions and answers

Dale Simms
January 17, 2025

I **** interested in your recommendation of cleaning products and disinfecting types. Currently we are using a wipe with 70% alcohol the a cavi wipe afterwards. We have had trouble with our glucometer shutting down, possibly from too much moisture. Also we start an IV and I would like to take the dot of blood from the IV angio- would this be acceptable with this glucometer? Thank you

1 comments:
Mr. Anderson
May 14, 2025

The recommended cleaning and disinfecting method for the McKesson PRO glucometer involves using wipes that have been tested for up to 10,950 cleaning and disinfecting cycles, equal to 10 cleanings per day over 3 years. This ensures the meter's proper function without moisture issues when done according to guidelines. Healthcare professionals must wear gloves while cleaning and disinfecting.

It is not acceptable to take a blood sample from an IV angio with this glucometer. The meter is validated for fresh capillary whole blood from the finger and forearm, and venous blood drawn in sodium heparin tubes—not from IV lines.

This answer is automatically generated

Anne Clark
March 7, 2025

how long after opening vial must control solution be discarded

1 comments:
Mr. Anderson
March 7, 2025

The control solution for the McKesson TRUE METRIX® PRO can be used for up to 3 months after first opening or until the expiration date printed on the label has passed, whichever comes first.

This answer is automatically generated

Sheila Sturm
March 19, 2025

what is a e-3 code

1 comments:
Mr. Anderson
May 14, 2025

An E-3 code for McKesson TRUE METRIX PRO indicates that the meter has detected the test strips are compromised.

This answer is automatically generated

Subscribe to Our Youtube Channel

Table of Contents

Save PDF