System Specifications
Result Range: 20-600 mg/dL
Sample Size: 0.5 microliter (0.5 µL)
Sample:Fresh capillary whole blood from the finger and
forearm, venous blood drawn in sodium heparin
tubes, or control solution
Test Time: Results in as little as 4 seconds
Result Value: Plasma values
Assay Method: Amperometric
Power Supply: One 3V lithium battery #CR2032
Battery Life: Approximately 1000 tests or 1 year
Automatic shut-off: After two minutes of non-use
Weight: 1.66 ounces
Size: 3.44" x 2.16" x 0.69"
Memory Size: 500 glucose results
Operating Range
Relative Humidity: 10-90% (Non-condensing)
Temperature: 41°F-104°F
Hematocrit: 20-70%
Altitude: Up to and including 10,200 feet.
Note: Use within specified environmental conditions only.
Chemical Composition
McKesson TRUE METRIX® PRO Professional Monitoring
Blood Glucose Test Strips: Glucose dehydrogenase-FAD
(Aspergillus species), mediators, buffers and stabilizers.
McKesson TRUE METRIX® Control Solution: Water,
d-glucose, buffers, viscosity enhancing agent, salts, dye and
preservatives.
(non-rechargeable)
(meter & test strips):