IMPORTANT SAFETY INFORMATION:

For the most accurate results using McKesson TRUE
METRIX® PRO:
• Read all product instructions for use before testing.
• Use of McKesson TRUE METRIX® PRO in a manner not
specified in this Owner's Manual is not recommended and
may affect ability to determine true blood glucose levels.
• McKesson TRUE METRIX® PRO is an in vitro (outside the
body) quantitative system for multiple patient use testing
of human whole blood only.
• Do not use for the diagnosis of or screening for
diabetes mellitus or for measuring blood glucose in
newborns.
• ONLY use auto-disabling single use lancing devices/
lancets to obtain capillary blood sample.
• Alternative site (forearm) testing should not be used to
calibrate continuous glucose monitors (CGMs).
• Alternative site (forearm) testing should not be used for
insulin dose calculations.
• DO NOT use venous whole blood collected in sodium
fluoride (grey top), lithium heparin, or EDTA vacutainer
tubes for testing, as this may cause inaccurate results.
• Use ONLY McKesson TRUE METRIX® PRO Professional
Monitoring Blood Glucose Test Strips and McKesson
TRUE METRIX® Control Solution with the McKesson
TRUE METRIX® PRO Professional Monitoring Blood
Glucose Meter.
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TRUE METRIX PRO and is the answer not in the manual?

Questions and answers

Dale Simms
January 17, 2025

I **** interested in your recommendation of cleaning products and disinfecting types. Currently we are using a wipe with 70% alcohol the a cavi wipe afterwards. We have had trouble with our glucometer shutting down, possibly from too much moisture. Also we start an IV and I would like to take the dot of blood from the IV angio- would this be acceptable with this glucometer? Thank you

1 comments:
Mr. Anderson
May 14, 2025

The recommended cleaning and disinfecting method for the McKesson PRO glucometer involves using wipes that have been tested for up to 10,950 cleaning and disinfecting cycles, equal to 10 cleanings per day over 3 years. This ensures the meter's proper function without moisture issues when done according to guidelines. Healthcare professionals must wear gloves while cleaning and disinfecting.

It is not acceptable to take a blood sample from an IV angio with this glucometer. The meter is validated for fresh capillary whole blood from the finger and forearm, and venous blood drawn in sodium heparin tubes—not from IV lines.

This answer is automatically generated

Anne Clark
March 7, 2025

how long after opening vial must control solution be discarded

1 comments:
Mr. Anderson
March 7, 2025

The control solution for the McKesson TRUE METRIX® PRO can be used for up to 3 months after first opening or until the expiration date printed on the label has passed, whichever comes first.

This answer is automatically generated

Sheila Sturm
March 19, 2025

what is a e-3 code

1 comments:
Mr. Anderson
May 14, 2025

An E-3 code for McKesson TRUE METRIX PRO indicates that the meter has detected the test strips are compromised.

This answer is automatically generated