Test Result (-P Only) - Fluke 9190A Operator's Manual

Ultra-cool drywell field metrology well
Table of Contents

Advertisement

9190A
Operators Manual
Item
DIRECTION controls whether the SETPOINTS are sequenced in one direction "→" (ascending
from 1 to 8) or in both directions "  " (ascending from 1 to 8 and then descending from 8 to 1)
before the sequence is repeated. If the both directions option is selected, the Program sequences
from the first SETPOINT to the last and then reverses direction sequencing from the last to the
first.
PASS TOLERANCE
test points that have large errors.
SETPOINTS MENU opens a menu to set each of the SETPOINTS for the Program. Only the
number of SETPOINTS defined by NO. SETPOINTS will be displayed.

Test Result (-P Only)

Table 2-14 shows and describes the menu selections on the Test Result menu. This menu
is available on the -P Model only.
MENU
Item
VIEW TEST
PRINT TEST
printed. The data is transmitted from the RS-232 or USB port in ASCII format. Terminal emulator
or other software can be used to receive the data and save it to a text file on a computer.
ERASE TESTS
2-16
Table 2-13. Ramp/Soak Setup Menu (cont.)
[-P Only]
is the allowable tolerance condition for the test and is used to highlight
Table 2-14. Test Result Menu
P R O G
T E S T
MENU
R E S U L T
[-P Only]
opens a menu where test results can be viewed.
[-P Only]
opens a menu to select a programs results and transmit them to a PC to be
[-P Only]
opens a menu to erase test results.
Function
66.03°C
T E S T R E S U L T M E N U
V I E W
P R I N T
T E S T
T E S T
1
Function
E R A S E
T E S T
2
3
gzs012.eps