Hach 2100P Instrument And Procedure Manual page 53

Portable turbidimeter
Table of Contents

Advertisement

SECTION 3,
READ
16.
When all the digits
show the appropriate
value, press
READ
The instrument will
count from 60 to 0
(67 to 0 if signal average
is on), measure the
turbidity and store the
value. The instrument
will increment back to
S0. Remove the sample
cell from the cell
compartment.
continued
17.
Press:
The instrument will store
.
the new calibration data
and return the instrument
to the measurement
mode. It will use the new
calibration to calculate
turbidity for subsequent
measurements.
Note: Pressing
completes the calculation
of the calibration
coefficients. If calibration
errors occurred during
calibration, error messages
will appear after
pressed. If
appear, check the standard
preparation and review the
calibration; repeat the
calibration if necessary.
CAL?
If
may have occurred during
calibration. If
flashing, the instrument
is using the default
calibration.
CAL
.
CAL
CAL
is
CAL
E 1
or
E 2
appears, an error
CAL?
is
53