Carl Zeiss
3.3.4
Insertion of Accessories for Incubation
• A temperature sensor (Fig. 23/1) can be used in the sample chamber body.
− Hold the temperature sensor so that it shows you the letter "J". Slide the temperature sensor
sleeve (Fig. 23/3) onto the top horizontal part of the J-shaped temperature sensor.
− Dip the temperature sensor with sleeve into the sample chamber body and position the
temperature sensor sleeve securely into the sample chamber body groove (Fig. 23/2).
1
Temperature sensor
2
Sample chamber body groove
3
Temperature sensor sleeve
Fig. 23
Insertion of temperature sensor
− Place a cover on top of the sample chamber body (Fig. 24/3). This will minimize evaporation of
medium or contamination of the sample chamber. If CO
(configuration dependent, chapter 2.7.2) the cover is recommended to maintain the CO
concentrations.
−
Keep the opening as small as possible. Use the 3 mm opening cover for capillaries (Fig. 24/1) and
7 mm opening cover for syringes (Fig. 24/2). The cover only rests on top of the sample chamber
body and is free to move.
30
CHAPTER 1 - HARDWARE
User Interfaces
000000-1790-528
Lightsheet Z.1
module Lightsheet Z.1 is used
2
gas
2
02/2013