Oxi 315i
Adjusting
AutoRange
AutoRead
Calibration
D. O. partial pressure
Measured parameter
Measured value
Measuring system
OxiCal
Oxygen saturation
Reset
Resolution
Salinity
Salt content
Sample
Slope
Glossary
To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains with-
in the tolerance.
Name of the automatic selection of the measuring range.
WTW name for a function to check the stability of the measured value.
Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Of-
ten, this expression is also used when the measuring system is adjust-
ed at the same time (see adjusting).
Pressure caused by the oxygen in a gas mixture or liquid.
The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
The measuring system comprises all the devices used for measuring,
e. g. measuring instrument and probe. In addition, there is the cable
and possibly an amplifier, terminal strip and armature.
®
WTW name for a procedure to calibrate D. O. measuring systems in
water vapor saturated air.
Short name for the relative D. O. saturation.
Note: The D. O. saturation value of air-saturated water and the D. O.
saturation value of oxygen-saturated water are different.
Restoring the original condition of all settings of a measuring system.
Smallest difference between two measured values that can be dis-
played by a measuring instrument.
The absolute salinity S
of the mass of dissolved salts to the mass of the solution (in g/Kg). In
practice, this dimension cannot be measured directly. Therefore, the
practical salinity is used for oceanographic monitoring. It is deter-
mined by measuring the electrical conductivity.
General designation for the quantity of salt dissolved in water.
Designation of the sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
The slope of a linear calibration function.
of seawater corresponds to the relationship
A
Lists
33