4.3.1 Starting the calibration

Proceed as follows to calibrate the instrument:
1
Connect the D. O. probe to the measuring instrument.
2
Keep the OxiCal
Note
The sponge in the air calibration vessel must be moist (not wet). Follow
the instructions in the OxiCal
3
Put the D. O. probe into the air calibration vessel.
Press the <CAL> key. The calibration mode appears.
4
TDS UpH
O
1
C
A L
S
8
1
7
Time
Baud
Day.Month No.
Year
Ident
Tref25 Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
5
Press the <RUN/ENTER> key. AutoRead is active, AR flashes.
TDS UpH
O
0
8 8
1
S
1
8
7
Time
Baud
Day.Month No.
Year
Ident
Tref25 Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
6
As soon as a stable value is reached, the AR display indicator
stops flashing. This indicates that the calibration is complete;
the relative slope that was determined and the sensor evalua-
tion appear on the display (see page 19).
®
-SL air calibration vessel ready.
®
-SL operating manual.
Sal
mV/pH
% mg/l
m
S/
cm
mbar
M
cm
1/ cm
Sal
6
°
C
/
%
K
TP
°
F
Auto Store
ARng
AR
RCL
Sal
mV/pH
% mg/l
S/
m
cm
mbar
M
cm
1/ cm
Sal
6
°
C
/
%
K
TP
°
F
Auto Store
ARng
AR
RCL
Oxi 315i

Advertisement

Table of Contents
loading

Table of Contents