Chapter 6: Repair and Replacement Procedures
NOTE: When re-installing the sensor housings on to the recorder assembly using the two socket-head screws,
make sure that the top of the sensor housing is aligned with the top of the print head bracket (See
Figure 6-54) before tightening the screws.
NOTE: When re-installing paper-low and paper-out sensor cable connectors to their counterpart connectors
on the recorder assembly, make sure the pin 1 of the cable connector (pin 1 is marked on the connector)
plugs into the pin 1 of its counterpart connector (pin 1 is marked on the top of the recorder assembly as
illustrated below).
Figure 6-51
Figure 6-53
112
2036947-001
Figure 6-52
Figure 6-54
© 2013 by General Electric Company. All rights reserved.