Summary of Contents for A&D HL-1000WP

  • Page 1 Digital Scale HL-WP Series HL-1000WP HL-300WP HL-3000WP WM+PD4000754...
  • Page 2: Safety Precautions

    SAFETY PRECAUTIONS All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute: Product Safety Signs and Labels). The meanings are as follows: A potentially hazardous situation which, if not WARNING avoided, could result in death or serious injury. A potentially hazardous situation which, if not CAUTION avoided, may result in minor or moderate injury.
  • Page 3: Parts Description

    Spirit level UNITS key (g/lb and g/lb-oz models only) Changes w eighing units, Four leveling feet g or lb (oz). g/lb model: HL-1000WP g/lb-oz model: HL-300WP/HL-3000WP Rubber cap Battery cover AC adapter jack (Batteries are not included.) Battery compartment STABLE indicator Indicates when the reading 01200.
  • Page 4 PREPARATION Installing the Batteries / Connecting The AC adapter Installing the Batteries Stop up with the rubber cap CAUTION When replacing batteries, be careful not to apply too much force to the weighing pan. An excessive force could damage the scale. Remove the battery cover and insert six batteries (R6P / LR6 / AA size) into the battery compartment, taking extreme...
  • Page 5: Before Weighing

    To power up with “g” selected: Turn the scale off. Press and hold the [UNITS] key, then press the [ON/OFF] key. The scale will start with “g”. To return to the original setting repeat this instruction. Note: g/lb model: HL-1000WP g/lb-oz model: HL-300WP/HL-3000WP...
  • Page 6: When To Calibrate

    Calibration using a calibration weight Calibrate the scale with calibration weights at full capacity. HL-1000WP: 1000g ± 0.1g HL-300WP: 300g ± 0.02g HL-3000WP: 3000g ± 0.2g 1. Enter calibration mode. Press the [ON/OFF] key to turn the power off.
  • Page 7 Calibration by gravity compensation If you have calibration weights at the capacity of the scale, you can calibrate the scale with the weights instead of compensating for cceleration due to gravity. The scale is calibrated at 9.798 m/sec at the factory. If the gravity acceleration at your location is not same, calibrate the scale by compensating for the gravity acceleration.
  • Page 8: Specifications

    SPECIFICATIONS MODEL HL-1000WP HL-300WP HL-3000WP Capacity 1000 g 300 g 3000 g 2.2 lb 10.58 oz 6 lb 9.8 oz Resolution 0.5 g 0.1 g 0.001 lb 0.01 oz 0.1 oz Non-linearity ±1 g ±0.2 g ±2 g Repeatability 0.5 g 0.1 g...
  • Page 9: World Map

    World Map...
  • Page 10 3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, INC 1555 McCandless Drive, Milpitas, CA. 95035 U.S.A. Telephone: [1] (408) 263-5333 Fax: [1] (408) 263-0119 A&D INSTRUMENTS LTD. Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxon, OX14 1DY United Kingdom.