Setting The Function Table; Details Of The Function Table - A&D HR-300i Instruction Manual

Hr-i series analytical balances
Table of Contents

Advertisement

Details of the Function Table

10.2.
Class
Item and Parameter
Cond
Condition
5t-b
Stability band width
ba5fnc
trc
Environment
Zero tracking
Display
5pd
Display refresh rate
pnt
Decimal point
p-on
Auto display-ON
prt
Data output mode
ap-p
Auto print polarity
ap-b
dout
Auto print difference
Data output
5-id
ID number output
pU5e
Data output pause
at-f
Auto feed
info
GLP output
ar-d
Zero after output
#1 The unit of minimum display is digit. Example: If 1 mg display is selected using the
key for the HR-300i, 1 mg is one digit.
#2 Usable minimum display of the balance is one digit. Example: In gram display, one digit is
0.1 mg for the HR-300i , and 0.01 mg for the HR-202i.
0
Fast response, sensitive value
1
2
Slow response, stable value
0
Stable range is ±1 digit
1
2
Stable range is ±3 digits
0
OFF
1
Normal
2
Strong
3
Very strong
0
5 times/second
1
10 times/second
0
Point (.)
1
Comma (,)
0
OFF
1
ON
0
Key mode
Auto print mode A
1
(Reference = zero)
Auto print mode B
2
(Reference = last stable value)
3
Stream mode
0
Plus only
1
Minus only
2
Both
0
10 digits
1
100 digits
2
1000 digits
No output
0
1
Output
0
No pause
1
Pause (1.6 seconds)
0
Not used
1
Used
0
No output
1
AD-8121 format
2
General data format
0
Not used
1
Used
: Factory settings.
28
Description
FAST
Can be changed by response
MID.
adjustment.
SLOW
The stabilization indicator illuminates
when the display fluctuation is within
#1
the range per second.
Keeps zero display by tracking zero
drift.
Period to refresh the display
Decimal point format
Turns on the weighing mode display
when the AC adapter is connected.
Accepts the PRINT key only when
the display is stable.
Outputs data when the display is
stable and conditions of ap-p, ap-b
and the reference value are met.
Outputs data continuously.
Displayed value>Reference
Displayed value<Reference
Regardless of displayed value
Difference between reference value
and displayed value
Selects whether or not the ID
number is output.
Selects the data output interval.
Selects whether or not automatic
feed is performed.
Selects GLP output method.
For how to set time and date to be
added, refer to peripheral manual.
Adjusts zero automatically after data
is output
#2
RANGE